Modeling scattered radiation from multi-leaf collimators (MLCs) to improve calculation accuracy of in-air output
So-Yeon Park1,2, Siyong Kim3, Wonmo Sung4
1Institute of Radiation Medicine, Seoul National University Medical Research Center, Seoul, Republic of Korea.
Abstract:
This study aims to model an extra-focal source for the scattered radiation from multi-leaf collimators (MLCs), namely an MLC scatter source, and to correct in-air output ratio (Sc) calculated using the conventional dual source model (DSM) to achieve better accuracy of point dose calculation. To develop the MLC scatter source, a 6 MV photon beam from a Varian Clinac® iX linear accelerator with millennium 120 MLCs was used. It was assumed that the position for the MLC scatter source was located at the center of the MLC, consisting of line-based and area-based sources to consider the characteristics of the scattered radiation from the MLCs empirically. Based on the measured Sc values for MLC-defined fields, optimal parameters for the line-based and area-based sources were calculated using optimization process. For evaluation of proposed method, measurements were taken for various MLC-defined square and irregular fields. The Sc values calculated using the proposed MLC scatter source and conventional DSM were compared with the measured data. For MLC-defined square fields, the measured Sc values showed better agreement with those calculated using the MLC scatter source (the mean difference was - 0.03%) compared with those calculated using the DSM (the mean difference was 0.18%). For MLC-defined irregular fields, the maximum dose differences between measurements and calculations using the MLC scatter source and DSM were 0.54% and 1.45%, respectively. The developed MLC scatter source could improve the accuracy of Sc calculation for both square and irregular fields defined by MLCs.
More Related Videos
06:46Author Spotlight: Enhanced Urodynamic Method for Precise Urine Measurement in Awake Mice with Neurogenic Bladder
Published on: June 7, 2024
10:14Author Spotlight: Leaf Trait Analysis for Climate and Ecology Reconstruction in Modern and Ancient Plant Communities
Published on: October 25, 2024
Related Concept Videos
Improving Translational Accuracy
Improving Translational Accuracy
Biological Effects of Radiation
Uncertainty in Measurement: Accuracy and Precision
Cardiac Output II: Effect of Stroke Volume on Cardiac Output
Preload
Preload refers to the initial elongation of the cardiac myocytes before contraction and is related to the volume of blood filling the heart at the end of diastole, or end-diastolic volume. The...
Cardiac Output I:Effect of Heart Rate on Cardiac Output
Cardiac output (CO) refers to the total amount of blood ejected by one of the ventricles in liters per minute (L/min). In a resting adult, CO ranges from 5 to 6 L/min, adjusting according to the body's metabolic requirements.
Effect of Heart Rate on Cardiac Output
Cardiac output adapts to metabolic demands during stress, physical activity, or illness. The autonomic nervous system regulates heart rate via the sinoatrial node. The parasympathetic nervous system decreases heart...