Vertical Off-Centering in Reduced Dose Chest-CT: Impact on Effective Dose and Image Noise Values
Matthias Eberhard1, Christian Blüthgen1, Borna K Barth1
1Institute of Diagnostic and Interventional Radiology, University Hospital Zurich, Switzerland.
Objectives:
To assess the effect of vertical off-centering in tube current modulation (TCM) on effective-dose and image-noise in reduced-dose (RD) chest-CT.
Methods:
One-hundred consecutive patients (36 female; mean age 56 years) were scanned on a 192-slice CT scanner with a standard-dose (ND) and a RD chest-CT protocol using tube current modulation. Image-noise was evaluated by placing circular regions of interest in the apical, middle, and lower lung regions. Two independent readers evaluated image quality. Study population was stratified according to patient position in the gantry: positioned in the gantry isocenter (i), higher than the gantry isocenter (ii), and lower than the gantry isocenter, (iii). Pearson correlation was used to determine the correlation between effective radiation dose and vertical off-centering. Student's t test was used to evaluate for differences in image-noise between groups (i-iii).
Results:
Mean vertical off-centering was of 10.6 mm below the gantry-isocenter (range -45.0-27.9 mm). Effective radiation dose varied in a linear trend, with the highest doses noted below gantry isocenter, and the lowest doses noted above gantry isocenter (ND: r = -0.296; p = 0.003 - RD: r = -0.258; p = 0.010). Lowest image-noise was observed where patients were positioned below the gantry isocenter, and highest in patients positioned above (ND: 79.35 HU vs. 94.86 HU - RD: 143.44 HU vs. 160.13 HU). Subjective image quality was not significantly affected by patient-position (p > 0.05). Overall, there was no over-proportional noise-increase from the ND to the RD protocol in patients which were positioned off-center.
Conclusion:
Vertical off-centering influences effective radiation dose and image-noise on ND and RD protocols.
Advances In Knowledge:
There is no over-proportional noise increase in RD compared to ND protocols when patients are positioned off-center.
Related Concept Videos
Insulin: Dosing Regimen and Adverse Effects
The basal dose constitutes about 40%-50% of the total daily dose, with the rest as premeal insulin. The mealtime insulin dose should mirror...
Dose Size and Dosing Frequency: Determination Methods
Determination of Multiple Dosing Parameters: Loading and Maintenance Doses
Rational Dosage Regimen: Maintenance Dose and Loading Dose
In most cases, drugs are administered repetitively or infused continuously to maintain a steady-state concentration in the body. At a steady...
Dose-Response Relationship: Overview
Drug Dosing: Infants and Children