Measuring the Mechanical Properties of Living Cells Using Atomic Force Microscopy
Published on: June 27, 2013
The comparison between force volume and peakforce quantitative nanomechanical mode of atomic force microscope in
Yang Yang1, Xiaoxiao Xiao1, Yan Peng1
1School of Mechatronics Engineering and Automation, Shanghai University, Shanghai, China.
Abstract:
Atomic force microscope (AFM) has been widely used in the biological field owing to its high sensitivity (subnanonewton), high spatial resolution (nanometer), and adaptability to physiological environments. Nowadays, force volume (FV) and peakforce quantitative nanomechanical (QNM) are two distinct modes of AFM used in biomechanical research. However, numerous studies have revealed an extremely confusing phenomenon that FV mode has a significant difference with QNM in determining the mechanical properties of the same samples. In this article, for the case of human benign prostatic hyperplasia cells (BPH) and two cancerous prostate cells with different grades of malignancy (PC3 and DU145), the differences were compared between FV and QNM modes in detecting mechanical properties. The results show measured Young's modulus of the same cells in FV mode was much lower than that obtained by QNM mode. Combining experimental results with working principles of two modes, it is indicated that surface adhesion is highly suspected to be a critical factor resulting in the measurement difference between two modes. To further confirm this conjecture, various weight ratios of polydimethylsiloxane (PDMS) were assessed by two modes, respectively. The results show that the difference of Young's modulus measured by two modes increases with the surface adhesion of PDMS, confirming that adhesion is one of the significant elements that lead to the measurement difference between FV and QNM modes.
More Related Videos
Related Concept Videos
Mechanical Protein Functions
Atomic Force Microscopy
The AFM Probe
The probe is regarded as the heart of any AFM setup and comprises the...