Related Experiment Video
Updated: Jan 20, 2026

Optimization of In vitro Transcription Reaction for mRNA Production Using Chromatographic At-Line Monitoring
Published on: April 4, 2025
Chromatographic studies of unusual on-column degradation of cefaclor observed in the impurity separation by HPLC
Jia-Sheng Zhong1, Yan Huang1, Jin-Zhi Wan2
1Research and Development Center, Sinopharm Zhijun (Shenzhen) Pharmaceutical Co., Ltd., Shenzhen 518110, PR China.
Abstract:
An unpredictable ghost peak was intermittently observed during the impurity separation of cefaclor and formulation by high performance liquid chromatography (HPLC) with a content from below the reported threshold to approximately 0.3% in different laboratories. Through a series of investigations, the ghost peak was identified as an unusual on-column degradant of cefaclor formed under elevated column temperature but was not an actual sample impurity. The chemical structure of the degradant was determined by spectroscopic methods, including high resolution mass spectrometry (HRMS) and 1H-NMR. Consequently, the unknown peak was identified as a C-4 oxidative decarboxylation analog of cefaclor. The formation mechanism of the analog is proposed, and it is suggested that elevated column temperature during HPLC analysis has a profound effect on the degradation. Dissolved oxygen in the mobile phase may promote the formation of the ghost peak. The degradation can be suppressed by using a column temperature below 30 °C. Moreover, several other prevention measures are suggested based upon the results of the investigation.
More Related Videos
12:04Thin-layer Chromatographic TLC Separations and Bioassays of Plant Extracts to Identify Antimicrobial Compounds
Published on: March 27, 2014
05:46Identification of Pharmaceuticals in The Aquatic Environment Using HPLC-ESI-Q-TOF-MS and Elimination of Erythromycin Through Photo-Induced Degradation
Published on: August 1, 2018
Related Concept Videos
Optimizing Chromatographic Separations
Band broadening refers to spreading solute bands as they travel through the column. This broadening can impact resolution. Plate height (H) represents the length required for one theoretical plate. A lower plate height corresponds to...
Unusual Results
According to the range rule of thumb, any value above or below two standard deviations, 2σ from the mean, μ is considered unusual.
Maximum unusual value =...
Observational Studies
There are three types of observational studies – Prospective, retrospective, and cross-sectional.
Prospective Study
Prospective studies, also known as longitudinal or cohort studies, are carried out by collecting future data from groups sharing similar characteristics. One...
z Scores and Unusual Values
This score indicates how far a value is from the mean in terms of standard deviation. For example, if a data value has a z score of +1, the researcher can infer that the particular data value is one standard deviation above the mean. If another data...
Chromatographic Resolution
The effectiveness of separation can be evaluated by determining the level of separation between two neighboring peaks in a chromatogram, which represents the individual components of a sample.
In chromatography,...
Chromatographic Methods: Terminology