Fabry-Pérot optical sensor and portable detector for monitoring high-resolution ocular hemodynamics
Jeong Oen Lee1, Vinayak Narasimhan2, Ashwin Balakrishna3
1California Institute of Technology, Pasadena, CA 91125 USA and now with the National Institute of Health, Bethesda, MD, 20852 USA (jeongoen.lee@nih.gov).
Abstract:
Our understanding of ocular hemodynamics and its role in ophthalmic disease progression remains unclear due to the shortcomings of precise and on-demand biomedical sensing technologies. Here, we report high-resolution in vivo assessment of ocular hemodynamics using a Fabry-Pérot cavity-based micro-optical sensor and a portable optical detector. The designed optical system is capable of measuring both static intraocular pressure and dynamic ocular pulsation profiles in parallel. Through a dynamic intensity variation analysis method which improves sensing resolution by 3-4 folds, our system is able to extract systolic/diastolic phases from a single ocular pulsation profile. Using a portable detector, we performed in vivo studies on rabbits and verified that ophthalmic parameters obtained from our optical system closely match with traditional techniques such as tonometry, electrocardiography, and photo-plethysmography.