Related Concept Videos

Gas Chromatography: Introduction01:13

Gas Chromatography: Introduction

3.5K
Gas chromatography (GC) is a technique for separating and analyzing volatile compounds in a sample. Its primary purpose is to identify and quantify components in complex mixtures, making it essential in fields such as environmental analysis, pharmaceuticals, and petrochemicals. GC is also called vapor-phase chromatography (VPC) or gas-liquid partition chromatography (GLPC).
In GC,  a sample is vaporized and mixed with an inert carrier gas (the mobile phase), which transports it through a...
3.5K
Gas Chromatography–Mass Spectrometry (GC–MS)01:14

Gas Chromatography–Mass Spectrometry (GC–MS)

6.3K
Gas chromatography–mass spectrometry (GC–MS) is the combination of analytical techniques of gas chromatography and mass spectrometry in a single instrument for analyzing a mixture of compounds. The gas chromatograph separates the compounds in the mixture, and the mass spectrometer analyzes each compound separately to determine the molecular masses and molecular structures.
A gas chromatograph consists of a long, narrow capillary column with a polysiloxane coating on the inner wall....
6.3K
Gas Chromatography: Types of Columns and Stationary Phases01:17

Gas Chromatography: Types of Columns and Stationary Phases

2.0K
Gas chromatography (GC) relies on stationary phases to separate and analyze components in a sample. There are two main types of stationary phases: liquid and solid. Liquid stationary phases are non-volatile, thermally stable, and chemically inert liquids coated onto the column. Solid stationary phases are particles of adsorbent material, such as silica gel or molecular sieves.
For an analyte to remain on the column for a sufficient amount of time, it must exhibit some level of compatibility (or...
2.0K
Optimizing Chromatographic Separations01:15

Optimizing Chromatographic Separations

794
Optimizing chromatographic separations is crucial for obtaining clean separations in a minimum amount of time. Optimization is required for several factors, including kinetic effects related to band broadening, plate height, capacity factor, and separation factor.
Band broadening refers to spreading solute bands as they travel through the column. This broadening can impact resolution. Plate height (H) represents the length required for one theoretical plate. A lower plate height corresponds to...
794
Gas Chromatography: Sample Injection Systems01:08

Gas Chromatography: Sample Injection Systems

1.2K
In gas chromatography, the sample is introduced as a vapor plug into the carrier gas stream for high efficiency and resolution. A microsyringe injects the sample solution into a heated sample port, vaporizing it and mixing it with the carrier gas. This process is important to ensure the sample is properly prepared for analysis. Thermally sensitive samples can be injected directly into the column and volatilized by slowly increasing the column temperature.
Two primary injection methods are used...
1.2K
Gas Chromatography: Overview of Detectors01:13

Gas Chromatography: Overview of Detectors

1.7K
Detectors in gas chromatography (GC) help identify and quantify the components of a mixture by translating chemical properties into measurable signals, which are displayed on a chromatogram. Detectors can be categorized into two main types: destructive and non-destructive.
A non-destructive detector allows a sample to be analyzed without altering or consuming it, meaning the sample can be collected after detection for further analysis. Examples include thermal conductivity detectors and...
1.7K

You might also read

Related Articles

Articles linked to this work by shared authors, journal, and citation graph.

Sort by
Same author

Two metrics for measuring orthogonality for two-dimensional chromatography.

Journal of chromatography. A·2018
Same author

Enhancing detectability of anabolic-steroid residues in bovine urine by actively modulated online comprehensive two-dimensional liquid chromatography - high-resolution mass spectrometry.

Analytica chimica acta·2018
Same author

Comprehensive lipidomic analysis of human plasma using multidimensional liquid- and gas-phase separations: Two-dimensional liquid chromatography-mass spectrometry vs. liquid chromatography-trapped-ion-mobility-mass spectrometry.

Journal of chromatography. A·2017
Same author

Size-exclusion chromatography using core-shell particles.

Journal of chromatography. A·2016
Same author

A procedure for comprehensive two-dimensional gas chromatography retention time locked dual detection.

Journal of chromatography. A·2016
Same author

Tunable secondary dimension selectivity in comprehensive two-dimensional gas chromatography.

Journal of chromatography. A·2013
Same journal

Targeting, Tracking, and Triggering: Versatile Applications of Zr-MOFs in Drug Delivery, Theranostic Platforms, and Advanced Biosensing.

Critical reviews in analytical chemistry·2026
Same journal

Fourier Transform Infrared Spectroscopy for Quantitative Determination of Drug Purity in Pharmaceutical Dosage Forms: Recent Advances, Validation Strategies, and Chemometric Applications.

Critical reviews in analytical chemistry·2026
Same journal

Characterization of Coffee Residues and Derived Biochar via Fourier-Transform Infrared Spectroscopy: Current Status and Outlook.

Critical reviews in analytical chemistry·2026
Same journal

Research Progress of Purely Organic Luminophores in the Field of Electrochemiluminescence.

Critical reviews in analytical chemistry·2026
Same journal

Recent Progress in Fluorescent Probes for Hydrazine Detection with Double-Bond Structural Features.

Critical reviews in analytical chemistry·2026
Same journal

Real-Sample Validation of Microchemical Sensors: Matrix Effects, Chemometric Reliability, and Field-Ready Analytical Performance.

Critical reviews in analytical chemistry·2026
See all related articles

Related Experiment Video

Updated: Dec 31, 2025

Chromatographic Fingerprinting by Template Matching for Data Collected by Comprehensive Two-Dimensional Gas Chromatography
10:14

Chromatographic Fingerprinting by Template Matching for Data Collected by Comprehensive Two-Dimensional Gas Chromatography

Published on: September 2, 2020

5.4K

Column Selection and Optimization for Comprehensive Two-Dimensional Gas Chromatography: A Review.

John Mommers1, Sjoerd van der Wal2

  • 1DSM Material Science Center, Geleen, The Netherlands.

Critical Reviews in Analytical Chemistry
|January 11, 2020
PubMed
Summary
This summary is machine-generated.

Comprehensive two-dimensional gas chromatography (GC×GC) method development is complex due to interdependent parameters and instrumental constraints. This review offers guidelines for optimizing GC×GC setup, column selection, and settings for improved separation efficiency.

Keywords:
two-dimensionalGas chromatographymethod development

More Related Videos

On-line Analysis of Nitrogen Containing Compounds in Complex Hydrocarbon Matrixes
07:49

On-line Analysis of Nitrogen Containing Compounds in Complex Hydrocarbon Matrixes

Published on: August 5, 2016

11.1K
Measurement of H2S in Crude Oil and Crude Oil Headspace Using Multidimensional Gas Chromatography, Deans Switching and Sulfur-selective Detection
08:37

Measurement of H2S in Crude Oil and Crude Oil Headspace Using Multidimensional Gas Chromatography, Deans Switching and Sulfur-selective Detection

Published on: December 10, 2015

19.8K

Related Experiment Videos

Last Updated: Dec 31, 2025

Chromatographic Fingerprinting by Template Matching for Data Collected by Comprehensive Two-Dimensional Gas Chromatography
10:14

Chromatographic Fingerprinting by Template Matching for Data Collected by Comprehensive Two-Dimensional Gas Chromatography

Published on: September 2, 2020

5.4K
On-line Analysis of Nitrogen Containing Compounds in Complex Hydrocarbon Matrixes
07:49

On-line Analysis of Nitrogen Containing Compounds in Complex Hydrocarbon Matrixes

Published on: August 5, 2016

11.1K
Measurement of H2S in Crude Oil and Crude Oil Headspace Using Multidimensional Gas Chromatography, Deans Switching and Sulfur-selective Detection
08:37

Measurement of H2S in Crude Oil and Crude Oil Headspace Using Multidimensional Gas Chromatography, Deans Switching and Sulfur-selective Detection

Published on: December 10, 2015

19.8K

Area of Science:

  • Analytical Chemistry
  • Chromatography

Background:

  • Comprehensive two-dimensional gas chromatography (GC×GC) offers enhanced separation power compared to 1D GC.
  • Method development in GC×GC is inherently complex due to numerous interdependent parameters and instrumental limitations.

Purpose of the Study:

  • To review existing literature on GC×GC method development.
  • To propose general guidelines for optimizing GC×GC instrumental setup, column selection, and operational parameters.

Main Methods:

  • Literature review of published GC×GC method development studies.
  • Analysis of common challenges and optimization strategies in GC×GC.

Main Results:

  • GC×GC method development involves complex interplay between primary and secondary column parameters.
  • Optimization is constrained by modulation criteria and column temperature limits.
  • Column diameter differences often lead to sub-optimal flow settings and reduced efficiency.

Conclusions:

  • Developing robust GC×GC methods requires careful consideration of instrumental configuration and column choices.
  • Adherence to proposed guidelines can facilitate improved separation performance and efficiency in GC×GC analyses.