Long-term ungrounded cable support for short-to-shield syndrome
Brian Ayers1, Christina Cheyne1, Katherine Wood1
1University of Rochester Medical Center Rochester NY USA.
Abstract:
Short-to-shield (STS) is a potential complication for left ventricular assist device (LVAD) patients supported by the HeartMate II (HMII) pump. This phenomenon occurs when a damaged internal wire within the driveline makes contact with the surrounding sheath, resulting in insufficient power delivery to the motor when connected to a grounded power base unit (PBU). An ungrounded cable can be used to negate these effects, but the long-term safety of this treatment strategy is unknown. In this case series, we present our institutional experience treating 17 STS patients with an ungrounded cable. In total, we present 4922 patient-days (13.4 patient-years) of ungrounded cable support after primary STS treatment. There were no deaths or complications related to STS. These data suggest that the long-term use of an ungrounded cable is a reasonable treatment option for patients who cannot or do not wish to undergo pump exchange or splice repair.
Related Concept Videos
Cable Subjected to Concentrated Loads
Cable Subjected to a Distributed Load
Cable Subjected to Its Own Weight
A generalized loading function is employed to analyze a cable subjected to its own weight. This function considers the force acting along the cable's arc length rather than its projected length, providing a more accurate...
Cable: Problem Solving
Equipotential Surfaces and Conductors
Three-Phase Short Circuit—Unloaded Synchronous Machine
This behavior occurs due to the magnetic flux produced by the short-circuit armature currents. Initially, these currents follow high-reluctance paths but eventually shift to...