A Novel Non-invasive Method for the Detection of Elevated Intra-compartmental Pressures of the Leg
Published on: May 31, 2019
Non-Invasive Central Venous Pressure Measurement Using Enclosed-Zone Central Venous Pressure (ezCVPTM)
Takayuki Hidaka1, Yoji Sumimoto1, Yoshihiro Dohi1
1Department of Cardiovascular Medicine, Hiroshima University Graduate School of Biomedical and Health Science.
Background:
Central venous pressure (CVP) is measured to assess intravascular fluid status. Although the clinical gold standard for evaluating CVP is invasive measurement using catheterization, the use of catheterization is limited in a clinical setting because of its invasiveness. We developed novel non-invasive technique, enclosed-zone (ezCVPTM) measurement for estimating CVP. The purpose of this study was to assess the feasibility of ezCVP and the relationship between ezCVP and CVP measured by a catheter.
Methods And Results:
We conducted 291 measurements in 97 patients. Linear regression analysis revealed that ezCVP was significantly correlated with CVP (r=0.65, P<0.0001). The Bland-Altman analysis showed that ezCVP had an underestimation bias of -2.5 mmHg with 95% limits of agreement of -14.1 mmHg and 9.6 mmHg for CVP (P<0.0001). The areas under the curves of receiver operating curve with ezCVP to detect the CVP ≥12 cmH2O (8.8 mmHg) and CVP >10 mmHg were 0.81 or 0.88, respectively. The sensitivity, specificity and positive likelihood ratio of ezCVP for the CVP ≥8.8 mmHg and CVP >10 mmHg were 0.59, 0.96 and 14.8 with a cut-off value of 11.9 and 0.79, 0.97 and 26.3 with a cut-off value of 12.7.
Conclusions:
These findings suggest that ezCVP measurement is feasible and useful for assessing CVP.
Related Concept Videos
Parentral Nutrition: Centeral and Peripheral Parental Nutrition
PN can be administered through two primary routes:
1. Central Parenteral Nutrition (CPN):
CPN involves delivering a high concentration of nutrients through a large vein. This is typically achieved using a Peripherally Inserted Central Catheter (PICC) or,...
Measurement of Blood Pressure
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Venous Thrombosis II: Clinical Manifestations and Diagnostic Studies
Sites for measruring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement