Buckling Resistance of Various Nickel-Titanium Glide Path Preparation Instruments in Dynamic or Static Mode
Jung-Hong Ha1, Sang Won Kwak2, Antheunis Versluis3
1Department of Conservative Dentistry, School of Dentistry, Kyungpook National University, Daegu, Korea.
Introduction:
The aim of this study was to compare the buckling resistance of nickel-titanium (NiTi) instruments for glide path preparation depending on the test mode (static vs dynamic).
Methods:
The conventional PathFile (PF; Dentsply Sirona, Ballaigues, Switzerland) and heat-treated ProGlider (PG, Dentsply Sirona) and WaveOne Gold Glider (WG, Dentsply Sirona) were evaluated. The instrument tips were placed in a small dimple prepared in an aluminum cube in a customized device. The file was then pushed in the axial direction at a 1.0-mm/s crosshead speed with rotation (dynamic mode) or without rotation (static mode). The dynamic mode of WG used its dedicated reciprocating rotation, whereas the others were rotated continuously at 300 rpm. The axial load and lateral buckling displacement were simultaneously measured. Data were analyzed statistically using 2-way analysis of variance (P = .05).
Results:
The buckling resistance in the dynamic mode was higher than in the static mode for PG and WG (P < .05), whereas PF was not influenced by test modes (P > .05). In the dynamic mode, the PG required the highest buckling load followed by PF and WG (P < .05). In the static mode, the WG showed the lowest load (P < .05). The dynamic mode showed significantly more upper level lateral buckling displacement than in the static mode (P < .05).
Conclusions:
When the glide path preparation instruments moved in the dynamic mode as in clinical situations, the buckling resistance of the heat-treated NiTi glide path instruments was higher than in the static condition. The heat-treated instruments may have better buckling resistance than the conventional NiTi instrument in clinical situations.