Comparison of physical image qualities and artifact indices for head computed tomography in the axial and helical
Ichiro Fujimura1, Katsuhiro Ichikawa2, Yohei Miura3
1Department of Radiological Technology, Rinku General Medical Center, 2-23, Rinku Orai-kita, Izumisano, Osaka, 598-8577, Japan. i-fujimura@rgmc.izumisano.osaka.jp.
Abstract:
This study aimed to validate the clinically demonstrated equivalency of the axial and helical scan modes (AS and HS, respectively) for head computed tomography (CT) using physical image quality measures and artifact indices (AIs). Two 64-row multi-detector row CT systems (CT-A and CT-B) were used for comparing AS and HSs with detector rows of 64 and 32. The modulation transfer function (MTF), noise power spectrum (NPS), and slice sensitivity profile were measured using a CT dose index corresponding to clinical use. The system performance function (SPF) was calculated as MTF2/NPS. The AI of streak artifacts in the skull base was measured using an image obtained of a head phantom, while the AI of motion artifacts was measured from images obtained during the head phantom was in motion. For CT-A, the 50%MTFs were 7% to 9% higher in the HS than the AS, and the higher MTFs of HS associated NPS increases. For CT-B, the MTFs and NPSs were almost equivalent between the AS and HS, respectively. Consequently, the SPFs of AS and HS were nearly identical for both CT systems. For both CT systems, the skull base AI did not differ significantly between AS and HS, while the motion AIs of HS were significantly better than of AS. The superior motion AI in the HS indicated the effectiveness of HS on moving patients.
More Related Videos
Related Concept Videos
Imaging Studies III: Computed Tomography
Imaging Studies for Cardiovascular System V: CT
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...