His bundle lead placement: Is His bundle captured?

Ronpichai Chokesuwattanaskul1, Krit Jongnarangsin1

  • 1Division of Cardiac Electrophysiology, University of Michigan Health Care, Ann Arbor, Michigan.

Insights

This study explores His bundle pacing in a patient with heart block. Analysis of intracardiac electrograms suggests His bundle capture may not have occurred, highlighting the complexity of pacing technologies.

Area of Science:

  • Cardiology
  • Electrophysiology
  • Medical Devices

Background:

  • A 59-year-old female received a dual-chamber pacemaker for intermittent complete heart block.
  • Baseline electrocardiogram revealed normal sinus rhythm with first-degree atrioventricular (AV) block and right bundle branch block.

Observation:

  • A His bundle lead was implanted, and intracardiac electrograms showed atrial-His and His-ventricular intervals of 186 ms and 110 ms, respectively.
  • Pacing from the His bundle lead with decreasing output did not result in significant QRS morphology changes.

Findings:

  • The study questions whether His bundle capture was achieved despite lead placement.
  • Intracardiac electrogram analysis revealed specific conduction intervals within the His Purkinje system.

Implications:

  • Understanding His Purkinje system physiology is crucial for advanced pacing techniques.
  • This case highlights the challenges and diagnostic considerations in His bundle pacing and conducting system pacing technologies.