Monofilament MgB2 Wire for a Whole-Body MRI Magnet: Superconducting Joints and Test Coils
Jiayin Ling1, John Voccio1, Youngjae Kim1
1Francis Bitter Magnet Laboratory, Massachusetts Institute of Technology, Cambridge, MA 02139 USA.
Abstract:
This paper presents recent results from our continued development of a 0.5 T whole-body MRI magnet at the Francis Bitter Magnet Laboratory. HyperTech Research Corp. (Columbus, OH) manufactures the MgB2 conductor for this project. During the past year, we have found that our technique, originally developed successfully to splice unreacted multifilament MgB2 wires, works much better, i.e., of higher reliability, with unreacted monofilament MgB2 wires. This has led us to wind the entire coil components in our persistent-mode MRI magnet with unreacted monofilament MgB2 wire, having a MgB2 core of 0.4 mm in diameter, an overall diameter of 0.8 mm bare, 1 mm S-glass insulated. To verify that these coils would not suffer from flux jumping, as they would if wound with monofilament NbTi wire, magnetization studies were performed on monofilament wires of MgB2 and NbTi (as a reference) at 4.2 K. For the monofilament MgB2 wire, the results were affirmative. To further ensure the absence of flux jumping that may quench these current-carrying coils, two test coils were wound with unreacted monofilament MgB2 wire. One MgB2 coil was operated in driven mode, while the other MgB2 coil, equipped with a persistent current switch and terminated with a superconducting joint, was operated in persistent mode. The operating temperature range was 4.2-15 K for these MgB2 coils. The driven mode coil was operated in self-field. The persistent mode coil achieved a persistent current of 100 A, corresponding to a self-field of ~1 T in the winding, for 1 hour with no measurable decay. Both test coils were operated quench free.
More Related Videos
11:27Magnetically-Assisted Remote Controlled Microcatheter Tip Deflection under Magnetic Resonance Imaging
Published on: April 4, 2013
10:07Simultaneous Recordings of Cortical Local Field Potentials, Electrocardiogram, Electromyogram, and Breathing Rhythm from a Freely Moving Rat
Published on: April 2, 2018
Related Concept Videos
Magnetic Field Due to Two Straight Wires
Magnetic Field Due To A Thin Straight Wire
Magnetic Force On Current-Carrying Wires: Example
Magnetic Resonance Imaging
Magnetic Field Of A Current Loop
Magnetic Fields
A magnetic field is defined by the force that a charged particle experiences...