Related Experiment Video
Updated: Dec 8, 2025

Author Spotlight: Investigating Early Events and Long-Term Effects of ACL Injuries for Osteoarthritis Progression
Published on: September 29, 2023
Accelerated MRI of the knee. Quality and efficiency of compressed sensing
Andra-Iza Iuga1, Nuran Abdullayev1, Kilian Weiss2
1Institute of Diagnostic and Interventional Radiology, Faculty of Medicine and University Hospital Cologne, University of Cologne, Kerpener Str. 62, 50937, Cologne, Germany.
Purpose:
To evaluate potential clinical acceleration factors of Compressed SENSE (CS)1 in direct comparison with SENSE for fat saturated (fs)2, proton density-weighted (PD)3 2D and 3D sequences of the knee.
Method:
Twenty healthy volunteers were scanned with a 3 T scanner, all receiving a standard, fs 2D PD, three CS (CS 2, CS 3, CS 5) as well as time-equivalent SENSE accelerations (S 2, S 3, S 5). The fs 3D PD sequence was acquired with four CS (CS 6, CS 8, CS 10, CS 15) and equivalent SENSE (S 5.72, S 7.69, S 9.57, S 14) factors. Three independent readers rated the images. Signal-to-noise, contrast-to-noise, root-mean-square error and structural similarity index were analyzed for objective evaluation.
Results:
Scan time decreased with increasing CS factor (2D CS 2: 145 s, 2D CS 3: 95 s, 2D CS 5: 57 s, 3D CS 6: 293 s, 3D CS 8: 220 s, 3D CS 10: 176 s, 3D CS 15: 119 s). The 2D standard sequence was rated best for diagnostic certainty and overall image impression with an average of 4.97 ± 0.10 and 4.80 ± 0.24 (all p < 0.05), except for 2D CS 2 and 2D S 2. For the 3D sequences, the standard sequence performed better for both parameters for CS 15, S 9.57 and S 4, as well as S 7.69 for overall image impression while CS 8 was non-inferior for all tested criteria and CS 10 only inferior for delineation of the anterior cruciate ligament, both outperforming the time-equivalent SENSE accelerations.
Conclusion:
Compressed SENSE can significantly decrease (34.39 % for 2D CS 2 and 54.17 % for 3D CS 10) scan time in knee imaging with unchanged diagnostic certainty and overall image impression compared to the clinical reference.
Related Concept Videos
Magnetic Resonance Imaging
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies III: Computed Tomography
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Radiological Investigation II: MRI and Ventilation Perfusion Scan
Magnetic Resonance Imaging (MRI) and Ventilation Perfusion Scans are two radiological investigations that offer detailed diagnostic images of the body, particularly lung structures.
MRI
MRI uses magnetic fields and radiofrequency signals to distinguish between normal and abnormal tissues. This technology provides a more detailed diagnostic image than CT scans, enabling it to characterize pulmonary nodules, stage bronchogenic carcinoma, and evaluate inflammatory activity in...
Imaging Studies for Cardiovascular System IV: CMRI