Rebound Tonometry over Soft Contact Lenses
Iva Krolo1, Boze Mihaljevic1, Aida Kasumovic1
1Department of Ophthalmology, University Hospital Center Sestre milosrdnice, Zagreb, Croatiaa.
Summary
Rebound tonometry accurately measures intraocular pressure (IOP) over contact lenses (CLs). IOP readings with CLs were reliable, showing good correlation with measurements without CLs.
Area of Science:
- Ophthalmology
- Optometry
- Biomedical Engineering
Background:
- Goldmann applanation tonometry (GAT) is the established gold standard for intraocular pressure (IOP) measurement.
- Accurate IOP measurement is critical for diagnosing and managing glaucoma and ocular hypertension.
Purpose of the Study:
- To evaluate the accuracy of rebound tonometry for measuring IOP while patients wear hydrogel and silicone hydrogel contact lenses (CLs).
- To assess the influence of different CL powers on IOP measurements using rebound tonometry.
Main Methods:
- 117 eyes of 61 habitual contact lens wearers (without prior ocular conditions) were included.
- Intraocular pressure was measured using a rebound tonometer (iCare) with and without hydrogel/silicone hydrogel CLs of varying powers (-9.50 to +10.00 D).
- Paired t-test and correlation analysis were used to compare measurements and assess reliability.
Main Results:
- A significant positive correlation (r = 0.59) was observed between IOP measurements with and without CLs.
- The mean IOP measured with CLs (20.74±5.19 mmHg) was slightly higher (1.95 mmHg difference, P <0.01) than without CLs (18.79±4.36 mmHg).
- Increased central corneal thickness (CCT) positively correlated with increased rebound tonometry measurement error (r = 0.43).
Conclusions:
- Rebound tonometry demonstrates good reliability for IOP measurement over various contact lenses.
- This method is feasible and accurate for clinical use in contact lens wearers.
- Further research may explore the impact of CL material and thickness on IOP measurement accuracy.
Related Concept Videos
Focusing of Light in the Eye
4.2K
Light rays enter the eye through the cornea, a transparent dome-shaped tissue that is the eye's outermost layer. The cornea bends or refracts, light rays traveling to the pupil. The shape of the cornea determines how much of the light is bent and whether the image will be focused correctly on the retina at the back of the eye. Once the light has passed through both refraction layers, it converges into a single focal point onto a small area. This is where photoreceptors start transforming...
4.2K
Open Angle Glaucoma: Treatment
832
In open-angle glaucoma, the iridocorneal angle remains open, but the trabecular meshwork becomes stiff, slowing down the outflow of aqueous humor. This causes a buildup of aqueous humor in the anterior chamber, leading to a sudden increase in intraocular pressure. The treatment for open-angle glaucoma focuses on reducing the elevated intraocular pressure by either decreasing the secretion of aqueous humor or increasing its outflow.
Drugs such as carbonic anhydrase inhibitors, α2- and...
Drugs such as carbonic anhydrase inhibitors, α2- and...
832
Angle Closure Glaucoma: Treatment
957
Angle-closure glaucoma, or closed-angle glaucoma, is an eye condition where the iris bulges out and blocks the iridocorneal angle, resulting in a buildup of aqueous humor and increased intraocular pressure. Immediate medical attention is necessary due to the sudden onset of symptoms. The treatment for angle-closure glaucoma includes short-term and long-term approaches. Short-term treatment involves using eye drops like pilocarpine to lower intraocular pressure by increasing aqueous humor...
957
Glaucoma: Overview
1.1K
Glaucoma is an eye condition characterized by increased intraocular pressure that damages the retina and optic nerve, leading to irreversible blindness if left untreated. The human eye has various components, including the cornea, iris, pupil, lens, and optic nerve. Aqueous humor is secreted by the epithelium of the ciliary body in the posterior chamber and flows through the trabecular meshwork and canal of Schlemm, maintaining normal intraocular pressure. The trabecular meshwork and the canal...
1.1K