An on-site dosimetry audit for high-energy electron beams
Leon de Prez1, Stan Heukelom2, Bartel Jansen1
1VSL - Dutch Metrology Institute, Delft, The Netherlands.
Background And Purpose:
External dosimetry audits are powerful quality assurance instruments for radiotherapy. The aim of this study was to implement an electron dosimetry audit based on a contemporary code of practice within the requirements for calibration laboratories performing proficiency tests. This involved the determination of suitable acceptance criteria based on thorough uncertainty analyses.
Materials And Methods:
Subject of the audit was the determination of absorbed dose to water, D w, and the beam quality specifier, R 50,dos. Fifteen electron beams were measured in four institutes according to the Belgian-Dutch code of practice for high-energy electron beams. The expanded uncertainty (k = 2) for the D w values was 3.6% for a Roos chamber calibrated in 60Co and 3.2% for a Roos chamber cross-calibrated against a Farmer chamber. The expanded uncertainty for the beam quality specifier, R 50,dos, was 0.14 cm. The audit acceptance levels were based on the expanded uncertainties for the comparison results and estimated to be 2.4%.
Results:
The audit was implemented and validated successfully. All D w audit results were satisfactory with differences in D w values mostly smaller than 0.5% and always smaller than 1%. Except for one, differences in R 50,dos were smaller than 0.2 cm and always smaller than 0.3 cm.
Conclusions:
An electron dosimetry audit based on absorbed dose to water and present-day requirements for calibration laboratories performing proficiency tests was successfully implemented. It proved international traceability of the participants value with an uncertainty better than 3.6% (k = 2).
More Related Videos
07:31Characterization of Recombination Effects in a Liquid Ionization Chamber Used for the Dosimetry of a Radiosurgical Accelerator
Published on: May 9, 2014
06:20Irradiator Commissioning and Dosimetry for Assessment of LQ α and β Parameters, Radiation Dosing Schema, and in vivo Dose Deposition
Published on: March 11, 2021
Related Concept Videos
Biological Effects of Radiation
Atomic Emission Spectroscopy: Overview
Scanning Electron Microscopy
Fundamental Principles
Accelerated...
Atomic Emission Spectroscopy: Lab