Characterization of Recombination Effects in a Liquid Ionization Chamber Used for the Dosimetry of a Radiosurgical Accelerator
Published on: May 9, 2014
The impact of ion chamber components onkfor reference dosimetry in MRgRT
Júlia Navarro Campos1, Jacco de Pooter2, Leon de Prez2
1Department of Radiotherapy, Leiden University Medical Center, The Netherlands.
Abstract:
For reference dosimetry in MRgRT,kis used to correct for the impact of the magnetic field on the chamber calibration coefficient. It has been demonstrated that for accurate simulation ofkthe dead volume (DV) must be considered. This work goes one step further by analysing the contribution of secondary electrons generated in the various chamber components tok. The Farmer-type chamber PTW 30013 geometry was modelled for two different DVs. Monte Carlo simulations were performed for a60Co source and a 7 MV MRI-linac and the model was validated against measurements. Both parallel (α = 0° or 180°) and perpendicular (α = 90° or 270°) orientations of the chamber and the magnetic (B) field were considered, and severalB-field strengths between 0 T and 1.5 T. To study the dose contribution to the reduced volume (RV = cavity - DV) from the secondary electrons produced in certain components of the chamber the labelling of the particles was implemented in the PENELOPE user code PENMAIN. A separate model with each solid component of the chamber modelled as liquid water was used to investigate the impact of material choice onk. Results show that simulatedkvalues agree better with the measuredkwhen the DV is considered. It is demonstrated that small components of the chamber impactkconsiderably, since the contribution to the RV-dose from the bodies closer to the RV is higher than withoutB. Moreover, it is seen that the impact to the dose in the RV is reduced when the material of each component is modelled as liquid water. Therefore, chamber design and, to a lesser extent, choice of material affectk, and an accurate geometrical model of the chamber components and its further validation are important for correct calculations ofk.
More Related Videos
Related Concept Videos
Biological Effects of Radiation
Magnetic Resonance Imaging