Related Experiment Video
Updated: Nov 8, 2025

Fabrication of Ti3C2 MXene Microelectrode Arrays for In Vivo Neural Recording
Published on: February 12, 2020
Wireless electrochemiluminescence at functionalised gold microparticles using 3D titanium electrode arrays
Samantha F Douman1, David Collins2, Loanda R Cumba2
1National Centre for Sensor Research, School of Chemical Sciences, Dublin City University, FutureNeuro SFI Research Centre, Dublin 9, Ireland. Robert.Forster@dcu.ie and SensorLab (UWC Sensor Laboratories), Chemical Sciences Building, University of Western Cape Town, Robert Sobukwe Road, Bellville 7535, Cape, South Africa.
Abstract:
Wireless electrochemiluminescence is generated using interdigitated, 3D printed, titanium arrays as feeder electrodes to shape the electric field. Gold microparticles (45 μm diameter), functionalised with 11-mercaptoundecanoic acid, act as micro-emitters to generate electrochemiluminescence from [Ru(bpy)3]2+, (bpy is 2,2'-bipyridine) where the co-reactant is tripropylamine. The oxide coated titanium allows intense electric fields, whose distribution depends on the geometry of the array, to be created in the absence of deliberately added electrolyte. COMSOL modelling and long exposure ECL imaging have been used to map the electric field distribution. Significantly, we demonstrate that by controlling the surface charge of the gold microparticles through the solution pH, the light intensity can be increased by a factor of more than 10.