Related Experiment Video
Updated: Nov 2, 2025

Use of a Percutaneous Ventricular Assist Device/Left Atrium to Femoral Artery Bypass System for Cardiogenic Shock
Published on: August 16, 2021
Continuous-Flow Left Ventricular Assist Device Explantation After More Than 5 Years of Circulatory Support and
George V Letsou1, Andrew C W Baldwin1, Andrew B Civitello1
1Center for Cardiac Support, Texas Heart Institute, Houston, Texas.
Abstract:
Continuous-flow left ventricular assist devices have proved to be effective, durable, life-saving tools in patients with end-stage heart failure. However, because of the risks associated with mechanical circulatory support (including stroke, infection, gastrointestinal bleeding, and device malfunction), the optimal goal of device therapy is myocardial recovery and device removal. Ventricular reconditioning and pump explantation after continuous-flow support have been reported; however, little is known about variables that govern the pace and degree of myocardial response in patients who experience such recovery. We describe our long-term pump-weaning strategy for a 25-year-old man who had a continuous-flow device implanted and then needed more than 5 years of support from it before developing cardiac reserve sufficient to enable pump explantation. To our knowledge, this is the longest period of uninterrupted continuous-flow device support to end in successful pump deactivation and a return to medical therapy. This case highlights the importance of actively and persistently pursuing a device-weaning strategy in all patients who need left ventricular assist device therapy.