Reduction of emission time for [68Ga]Ga-PSMA PET/CT using the digital biograph vision: a phantom study
Pedro Fragoso Costa1, Walter Jentzen2, Finja SÜßELBECK2
1Department of Nuclear Medicine, Essen University Hospital, Duisburg-Essen University, Essen, Germany - pedro.fragoso-costa@uni-due.de.
Background:
The aim of this phantom study was to optimize the [68Ga]Ga-PSMA PET/CT examination in terms of scan time duration and image reconstruction parameters, in combination with PSF and TOF modelling, in a digital Biograph Vision PET/CT scanner.
Methods:
Three types of phantoms were used: 1) soft-tissue tumor phantom consisting of six spheres mounted in a torso phantom; 2) bone-lung tumor phantom; 3) resolution phantom. Phantom inserts were filled with activity concentrations (ACs) that were derived from clinical data. Phantom data were acquired in list-mode at one bed position. Images with emission data ranging from 30 to 210 s in 30-s increments were reconstructed from a reference image acquired with 3.5-min emission. Iterative image reconstruction (OSEM), point-spread-function (PSF) and time-of-flight (TOF) options were applied using different iterations, Gaussian filters, and voxel sizes. The criteria for image quality was lesion detectability and lesion quantification, evaluated as contrast-to-noise ratio (CNR) and maximum AC (peak AC), respectively. A threshold value of CNR above 6 and percentage maximum AC (peak AC) deviation range of ±20% of the reference image were considered acceptable. The proposed single-bed scan time reduction was projected to a whole-body examination (patient validation scan) using the continuous-bed-motion mode.
Results:
Sphere and background ACs of 20 kBq/mL and 1 kBq/mL were selected, respectively. The optimized single-bed scan time was approximately 60 s using OSEM-TOF or OSEM-TOF+PSF (four iterations, 4.0-mm Gaussian filter and almost isotropic voxel size of 3.0-mm side length), resulting in a PET spatial resolution of 6.3 mm for OSEM-TOF and 5.5 mm for OSEM-TOF+PSF. In the patient validation, the maximum percentage difference in lesion quantification between standard and optimized protocol (whole-body scan time of 15 vs. 5 min) was below 19%.
Conclusions:
A reduction of single-bed and whole-body scan time for [68Ga]Ga-PSMA PET/CT compared to current recommended clinical acquisition protocols is postulated. Clinical studies are warranted to validate the applicability of this protocol.
More Related Videos
06:53Management of Respiratory Motion Artefacts in 18F-fluorodeoxyglucose Positron Emission Tomography using an Amplitude-Based Optimal Respiratory Gating Algorithm
Published on: July 23, 2020
08:36Multi-Tracer Studies of Brain Oxygen and Glucose Metabolism Using a Time-of-Flight Positron Emission Tomography-Computed Tomography Scanner
Published on: June 7, 2024
Related Concept Videos
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body...
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET