New algorithm for corneal densitometry assessment based on anterior segment optical coherence tomography.
Xiao Yu Wang1, Tian Qiao Zhang2, Anil Rahul Rachwani3
1Institute of Clinic Ophthalmology, Clinic Hospital of Barcelona, University of Barcelona, Barcelona, Spain. xwanwanw7@alumnes.ub.edu.
Eye (London, England)
|August 3, 2021
Summary
A new algorithm objectively measures corneal densitometry using swept source anterior segment ocular coherence tomography (SS-AS-OCT) images. This study establishes standard densitometry values for normal eyes, aiding clinical and research applications.
Area of Science:
- Ophthalmology
- Medical Imaging
- Biomedical Engineering
Background:
- Corneal densitometry is a crucial parameter in ophthalmology.
- Objective and standardized measurement methods are needed for accurate assessment.
- Swept source anterior segment ocular coherence tomography (SS-AS-OCT) offers high-resolution imaging of the cornea.
Purpose of the Study:
- To introduce a novel algorithm for quantifying corneal densitometry from SS-AS-OCT images.
- To establish normative corneal densitometry values in a healthy population.
- To validate the algorithm's capability for objective assessment.
Main Methods:
- A MATLAB-based algorithm was developed to segment the cornea into anterior, posterior, and mid-stroma layers.
- The cornea was analyzed in two concentric areas: 0-2 mm and 2-4 mm.
- Mean corneal densitometry was calculated in grayscale units (GSU) for 111 healthy participants (195 eyes).
Main Results:
- The mid-stroma layer showed the highest densitometry (87.4 ± 12.1 GSU), while the anterior layer had the lowest (81.9 ± 14.2 GSU).
- The central 0-2 mm area exhibited significantly higher densitometry (97.8 ± 12.7 GSU) than the 2-4 mm area.
- No significant correlation was found between corneal densitometry and participant gender or age.
Conclusions:
- The developed MATLAB algorithm provides an objective method for assessing corneal densitometry using SS-AS-OCT.
- This study presents standard, normative data for corneal densitometry, beneficial for clinical diagnosis and research.
- The findings support the utility of SS-AS-OCT in detailed corneal analysis.