Jove
Visualize
Contact Us
JoVE
x logofacebook logolinkedin logoyoutube logo
ABOUT JoVE
OverviewLeadershipBlogJoVE Help Center
AUTHORS
Publishing ProcessEditorial BoardScope & PoliciesPeer ReviewFAQSubmit
LIBRARIANS
TestimonialsSubscriptionsAccessResourcesLibrary Advisory BoardFAQ
RESEARCH
JoVE JournalMethods CollectionsJoVE Encyclopedia of ExperimentsArchive
EDUCATION
JoVE CoreJoVE BusinessJoVE Science EducationJoVE Lab ManualFaculty Resource CenterFaculty Site
Terms & Conditions of Use
Privacy Policy
Policies

Related Concept Videos

Spectrophotometry: Introduction01:16

Spectrophotometry: Introduction

5.5K
Spectrophotometry is the quantitative measurement of the absorption, reflection, diffraction, or transmission of electromagnetic radiation through a material as a function of the intensity and wavelength of the radiation. A spectrophotometer is a device used to measure the change in the radiation intensity caused by its interaction with the material.
The essential components of a spectrophotometer include a source of electromagnetic radiation, a slot for placing a material to be analyzed, and a...
5.5K
UV–Vis Spectrometers01:14

UV–Vis Spectrometers

1.9K
The absorbance of UV and visible (UV–visible) radiations is measured using a UV–visible spectrophotometer. Deuterium lamps, which emit UV radiation, and tungsten lamps, which produce radiation in the visible region, are used as light sources in UV–visible spectrophotometers. A monochromator or prism is used for diffraction grating, i.e., to split the incoming radiation into different wavelengths. A system of slits is used to focus the desired wavelength on the sample cell.
1.9K
Instrument Calibration01:12

Instrument Calibration

348
Instrument calibration is essential for ensuring that instruments produce accurate and consistent results. It is vital in manufacturing, healthcare, testing laboratories, and scientific research. Calibration processes are specific to each instrument and help enhance data accuracy. Each instrument has a unique calibration process tailored to its design and function to improve data accuracy.
Analytical Balance Calibration
An analytical balance measures mass and requires regular calibration to...
348
UV–Vis Spectroscopy: Beer–Lambert Law01:09

UV–Vis Spectroscopy: Beer–Lambert Law

5.0K
The Beer-Lambert law describes the relationship between absorbance and concentration, which combines the principles established by scientists Johann Heinrich Lambert and August Beer. Lambert's law states that when light passes through a medium, the loss in intensity is directly proportional to the original intensity and the path length of the light. Beer's law proposed that the transmittance of a solution remains constant if the product of concentration and path length is constant. The...
5.0K
UV–Vis Spectrum01:30

UV–Vis Spectrum

1.5K
When light passes through a substance, a portion of the light is absorbed while the remaining light is reflected or transmitted. If the molecule absorbs light between the wavelengths of 180–400 nm range, the UV spectrum is obtained, and if it absorbs light in the 400–780 nm wavelength range, the visible spectrum is obtained.     
The UV–Vis spectrum of a molecule is the plot of its absorbance versus wavelength. The plot is drawn by taking molar...
1.5K
High-Resolution Mass Spectrometry (HRMS)01:15

High-Resolution Mass Spectrometry (HRMS)

1.8K
The resolution of a mass spectrometer depends on the efficiency of separating ions with different ion masses. The mass of an atom is approximated to the sum of the masses of protons and neutrons inside, considering the masses of protons and neutrons as equal. However, the masses of the proton (1.6726 × 10−24 g) and neutron (1.6749 × 10−24 g) are not truly equal. There is a minor error in the expression of atomic masses relative to the simplest atom of hydrogen. For...
1.8K

You might also read

Related Articles

Articles linked to this work by shared authors, journal, and citation graph.

Sort by
Same author

A Study of the Polarization of Fluorescence of Ordered Systems With Application to Ordered Liquid Crystals.

Journal of research of the National Bureau of Standards. Section A, Physics and chemistry·2020
Same author

Reflection Correction for High-Accuracy Transmittance Measurements on Filter Glasses.

Journal of research of the National Bureau of Standards. Section A, Physics and chemistry·2020
Same author

Adaptation of a High-Accuracy Spectrophotometer for Ultraviolet Work.

Journal of research of the National Bureau of Standards. Section A, Physics and chemistry·2020
Same author

Polarization Effects on Fluorescence Measurements.

Journal of research of the National Bureau of Standards. Section A, Physics and chemistry·2020
Same author

Numerical Evaluation of Diffraction Integrals.

Journal of research of the National Institute of Standards and Technology·2016
Same author

Computation of Fresnel Integrals. II.

Journal of research of the National Institute of Standards and Technology·2016
Same journal

The Photochemistry of Propane at High Photon Energies (8.4-21.2 eV).

Journal of research of the National Bureau of Standards. Section A, Physics and chemistry·2021
Same journal

Isomerization Processes in Ions of the Empirical Formula <math> </math>.

Journal of research of the National Bureau of Standards. Section A, Physics and chemistry·2021
Same journal

Temperature Dependence of Photocurrents Produced by X and Gamma Rays in Silicon Radiation Detectors.

Journal of research of the National Bureau of Standards. Section A, Physics and chemistry·2021
Same journal

Stable Radical-Anions Derived from Glyoxal <i>Bis</i>(phenylhydrazones).

Journal of research of the National Bureau of Standards. Section A, Physics and chemistry·2021
Same journal

High-Speed (Subsecond) Measurement of Heat Capacity, Electrical Resistivity, and Thermal Radiation Properties of Niobium in the Range 1500 to 2700 K.

Journal of research of the National Bureau of Standards. Section A, Physics and chemistry·2021
Same journal

A New Determination of the Atomic Weight of Zinc.

Journal of research of the National Bureau of Standards. Section A, Physics and chemistry·2021
See all related articles

Related Experiment Video

Updated: Oct 19, 2025

Characterization of Biological Absorption Spectra Spanning the Visible to the Short-Wave Infrared
07:38

Characterization of Biological Absorption Spectra Spanning the Visible to the Short-Wave Infrared

Published on: January 10, 2025

2.0K

Physical Parameters in High-Accuracy Spectrophotometry.

K D Mielenz1

  • 1Optical Physics Division, Institute for Basic Standards, National Bureau of Standards, Washington, D.C. 20234.

Journal of Research of the National Bureau of Standards. Section A, Physics and Chemistry
|September 27, 2021
PubMed
Summary
This summary is machine-generated.

This study introduces a new spectrophotometer design that minimizes errors from beam geometry and polarization. It achieves high photometric precision, enabling accurate transmittance measurements for filters and liquid samples.

Keywords:
High accuracy spectrophotometryhigh accuracylinearity test of photodetectorphysical parametersspectrophotometry

More Related Videos

UV-Vis Spectroscopic Characterization of Nanomaterials in Aqueous Media
05:16

UV-Vis Spectroscopic Characterization of Nanomaterials in Aqueous Media

Published on: October 25, 2021

10.2K
A Technical Guide for Performing Spectroscopic Measurements on Metal-Organic Frameworks
10:13

A Technical Guide for Performing Spectroscopic Measurements on Metal-Organic Frameworks

Published on: April 28, 2023

2.7K

Related Experiment Videos

Last Updated: Oct 19, 2025

Characterization of Biological Absorption Spectra Spanning the Visible to the Short-Wave Infrared
07:38

Characterization of Biological Absorption Spectra Spanning the Visible to the Short-Wave Infrared

Published on: January 10, 2025

2.0K
UV-Vis Spectroscopic Characterization of Nanomaterials in Aqueous Media
05:16

UV-Vis Spectroscopic Characterization of Nanomaterials in Aqueous Media

Published on: October 25, 2021

10.2K
A Technical Guide for Performing Spectroscopic Measurements on Metal-Organic Frameworks
10:13

A Technical Guide for Performing Spectroscopic Measurements on Metal-Organic Frameworks

Published on: April 28, 2023

2.7K

Area of Science:

  • Spectrophotometry
  • Optical Physics
  • Analytical Chemistry

Background:

  • Apparent transmittance measurements are influenced by spectrophotometer beam geometry and light polarization.
  • Focused light beams lead to geometry-dependent transmittance, differing from true transmittance.
  • Stray light and interference effects can further compromise measurement accuracy.

Purpose of the Study:

  • To develop a new spectrophotometer design that overcomes limitations of traditional instruments.
  • To achieve high photometric precision for accurate transmittance measurements.
  • To improve methods for testing detector linearity and correcting for nonlinearity.

Main Methods:

  • Utilizing collimated incident light to eliminate beam geometry effects on transmittance.
  • Employing mirror optics to reduce stray light.
  • Optimizing monochromator slit width to minimize interference.
  • Developing a curve-fitting procedure for detector nonlinearity correction.

Main Results:

  • The new spectrophotometer design achieves shot-noise limited photometric precision, enabling measurements below 10^-4 transmittance units.
  • The proposed curve-fitting method offers improved precision for detector nonlinearity correction compared to conventional methods.
  • Data on detector nonlinearity and its wavelength dependence were presented.

Conclusions:

  • The developed spectrophotometer design effectively eliminates errors associated with beam geometry and polarization.
  • The instrument offers exceptional photometric precision, suitable for advanced optical and chemical analysis.
  • Improved methods for detector linearity testing and correction enhance overall measurement reliability.