Comparison Between an Intraocular Lens With Extended Depth of Focus (Tecnis Symfony ZXR00) and a New Monofocal
Young Joon Jeon1, Yisang Yoon1, Tae-Im Kim2
1Department of Opthalmology, Kim's Eye Hospital, Konyang University College of Medicine, Seoul, Republic of Korea.
Purpose:
This study compared the extended depth of focus (EDOF) intraocular lens (IOL) (ZXR00; Tecnis Symfony, Johnson & Johnson Vision, Santa Ana, CA, US) to a novel, higher-order aspheric monofocal IOL (ICB00; Tecnis Eyhance, Johnson & Johnson Vision, Santa Ana, CA, US) which uses the same platform and material.
Methods:
Medical records of patients undergoing cataract surgery with ZXR00 or ICB00 implantation between March 2020 and January 2021 and with the data available for the 3-month visit were reviewed. The uncorrected near, intermediate, and distance visual acuity (VA); corrected distance VA; and optical quality parameters were the main outcome measures.
Results:
Among the 174 enrolled patients, 72 and 102 received the ZXR00 and ICB00, respectively. The average patient ages were 59.6 ± 10.6 (range: 49 to 70) and 65.2 ± 8.2 (range: 45 to 82) years in the ZXR00 and ICB00 groups, respectively, with significantly older patients in the ICB00 group. The other baseline parameters were not different for the 2 groups. Compared to the ICB00 group, the ZXR00 group showed markedly superior near VA (P < 0.05) at 3 months postoperatively. In terms of optical quality, ICB00 was, statistically, significantly superior to ZXR00.
Conclusions:
The ZXR00 showed remarkable near vision and defocus curve smoothness, while the ICB00 achieved better optical quality. The 2 IOLs had comparable distance and intermediate vision.
More Related Videos
04:59Author Spotlight: Enhancing Visual Outcomes in Cataract Surgery: A Novel Technique to Prevent Posterior Capsular Opacification Through IOL Rotation
Published on: July 7, 2023
05:46Author Spotlight: Advancements in Refractive Surgical Correction for Presbyopia and Exploring Postoperative Visual Acuity
Published on: September 20, 2024
Related Concept Videos
Focusing of Light in the Eye
Phase Contrast and Differential Interference Contrast Microscopy
In-phase-contrast microscopes, interference between light directly passing through a cell and light refracted by cellular components is used to create high-contrast, high-resolution images without staining. It is the oldest and simplest type of microscope that creates an image by altering the wavelengths of light rays passing through the specimen. Altered wavelength paths are created using an annular stop in the condenser. The annular stop produces a hollow cone of...
Imaging Biological Samples with Optical Microscopy
In optical microscopy, the specimen to be viewed is placed on a glass slide and clipped on the stage...