Related Experiment Video
Updated: Oct 13, 2025

An Anesthesia, Surgery, and Harvest Method for the Evaluation of Transpedicular Screws Using an In Vivo Porcine Lumbar Spine Model
Published on: May 31, 2017
Optimizing Lumbar Pedicle Screw Trajectory Utilizing a 3D-Printed Drill Guide to Ensure Placement of Pedicle Screws
Xin Zhao1, Jie Zhao1, Xiao-Jiang Sun1
1Shanghai Key Laboratory of Orthopaedic Implants, Department of Orthopaedic Surgery, Shanghai Ninth People's Hospital, Shanghai Jiaotong University School of Medicine, Shanghai, China.
Background:
Lower preoperative Hounsfield Unit (HU) values of vertebral body are associated with pedicle screw (PS) loosening after implantation with traditional trans-pedicular trajectory. However, the relationship between trajectory HU value and PS fixation quality remains unknown. This study aimed to investigate if 3-dimensionally (3D)-printed guider directed accurate implantation of pedicle screw could increase the anti-pulling properties of screws.
Methods:
3D models of cadaveric spines were reconstructed by using computed tomography image and PS trajectories were designed for both sides of vertebra. The designed trajectories were divided into high HU group and low HU group. PS implantation with 3D-printed screw guide can be in complementary shape with target vertebra. Throughout 3D finite element analysis and biomechanical tests, the pull-out strength of screws in high or low trajectory HU groups were compared.
Results:
The HU value was 132 ± 13 (mean ± standard deviation) in low HU group and 189 ± 17 in high HU group. The distance between planned trajectories and actual trajectories was 1.69 ± 0.4 mm. Biomechanical tests showed that in the high trajectory HU group the pull-out strength of screws was 750.41 ± 80.65 N; compared with 655.83 ± 74.31 N in the low trajectory HU group, the difference was statistically significant. When simulated with the finite element method, the pull-out strength of low HU trajectory pedicle screws was lower than that of high HU trajectory pedicle screws.
Conclusions:
Preoperative computer-assisted trajectory design using a 3D-printed screw guide may direct more accurate implantation with optimal implantation trajectory, and may provide a new way to improve pedicle screw fixation.