Objectives:
The purpose of this study was to evaluate the dimensional stability of prototyped surgical guides after autoclave and 2% glutaraldehyde sterilization processes.
Material And Methods:
Twenty prototyped surgical guides were prepared and submitted to two sterilization processes (n = 10): Physics - sterilization by autoclave (saturated water vapor under pressure, temperature of 126 to 130 °C, pressure of 1,7 at 1.9 kgf/cm2, 16 min); Chemistry - sterilization by 2% glutaraldehyde for 10 h. Six pre-established points were measured in the prototyped surgical guides, before and after sterilization, using a digital caliper rule. The comparisons were made using orthogonal contrasts using the linear model of mixed effects (random and fixed).
Results:
there are no significant differences between autoclave and glutaraldehyde 2% (p>0.05) there are significant differences after autoclave sterilization (p<0.05) and there are not significant differences after 2% glutaraldehyde sterilization (p>0.05).
Conclusion:
autoclave promoted dimensional alteration of the prototyped surgical guides, and the chemical sterilization by glutaraldehyde 2% did not cause dimensional alteration of the prototype surgical guides, being a favorable choice for sterilization.
Clinical Significance:
the sterilization of surgical guides can be performed through the chemical process with 2% glutaraldehyde without changing the linear precision of the prototype surgical guides.
Related Concept Videos
Methods of Sterilization II: Chemical Methods
Using chemical sterilization rather than heat to clean out equipment is recommended. It eradicates and removes all bacteria,...
Methods of Sterilization I: Physical Methods
Steam sterilization uses non-toxic, low-cost moist heat in the form of saturated steam under pressure, which is fast, microbicidal, and sporicidal, and quickly warms and penetrates fabrics. Autoclaves, or steam sterilizers, expose each item to direct steam contact for a predetermined time at the necessary...
Cleaning, Sterilization, and Disinfection
Cleaning
The cleaning process usually involves using water with detergents or enzymatic cleaner and removing foreign material from objects and surfaces, including organic material such as body fluids or inorganic material like soil. Cleaning is performed before high-level disinfection and sterilization because foreign materials on the cover of the devices interfere with process...