Construction of a Preclinical Multimodality Phantom Using Tissue-mimicking Materials for Quality Assurance in Tumor Size Measurement
Published on: July 29, 2013
Comparison of acquisition and iterative reconstruction parameters in abdominal computed tomography-guided procedures:
Julien Frandon1, Philippe Akessoul1, Aymeric Hamard1
1Department of Medical Imaging, Nîmes University Hospital, University of Montpellier, Medical Imaging Group Nîmes, EA 2992, Nîmes, France.
Background:
Many computed tomography (CT) navigation systems have been developed to help radiologists improve the accuracy and safety of the procedure. We evaluated the accuracy of one CT computer-assisted guided procedure with different reduction dose protocols.
Methods:
A total of 128 punctures were randomly made by two operators on two different anthropomorphic phantoms. The tube voltage was fixed to 100 kVp. Tube currents (mAs) were defined to obtain 4 dose levels: 180 mAs (D1.00), 90 mAs (D0.50), 45 mAs (D0.25) and 15 mAs (D0.10) with respective volume CT dose index (CTDIvol) of 7.02, 3.52, 1.75 and 0.59 mGy. The raw data were reconstructed using level 2 of advanced model-based iterative reconstruction (ADMIRE) (A2) for D1.00, A3 for D0.50, A4 for D0.25 and A5 for D0.10. Two 12-mm targets per phantom were selected. The mean Euclidean distance (EuD) between the tip of the needle and the isocenter of the target was measured for each puncture. The different measures were compared by paired Student's t-tests.
Results:
The mean EuD was 7.0±3.1 mm for the 128 punctures performed. Regardless of which phantom was considered, no significant difference in accuracy occurred between the 4 dose levels, which were 7.1±3.5 mm for D1.00; 7.1±3.1 mm for D0.50; 7.2±3.0 mm for D0.25 and 6.6±2.6 mm for D0.10.
Conclusions:
Abdominal CT-guided procedures, using computer-assisted navigation and iterative reconstruction algorithms, allow precise punctures on anthropomorphic phantoms with a dose reduction of -92% compared to a standard protocol.
More Related Videos
08:41Patient-Specific Polyvinyl Alcohol Phantom Fabrication with Ultrasound and X-Ray Contrast for Brain Tumor Surgery Planning
Published on: July 14, 2020
09:57Development and Evaluation of 3D-Printed Cardiovascular Phantoms for Interventional Planning and Training
Published on: January 18, 2021
Related Concept Videos
Imaging Studies III: Computed Tomography
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...