Direct Conversion X-ray Detector with Micron-Scale Pixel Pitch for Edge-Illumination and Propagation-Based X-ray
Abdollah Pil-Ali1,2, Sahar Adnani1,2, Christopher C Scott3
1Department of Electrical and Computer Engineering, University of Waterloo, 200 University Ave W, Waterloo, ON N2L 3G1, Canada.
Abstract:
In this work, we investigate the potential of employing a direct conversion integration mode X-ray detector with micron-scale pixels in two different X-ray phase-contrast imaging (XPCi) configurations, propagation-based (PB) and edge illumination (EI). Both PB-XPCi and EI-XPCi implementations are evaluated through a wave optics model-numerically simulated in MATLAB-and are compared based on their contrast, edge-enhancement, visibility, and dose efficiency characteristics. The EI-XPCi configuration, in general, demonstrates higher performance compared to PB-XPCi, considering a setup with the same X-ray source and detector. However, absorption masks quality (thickness of X-ray absorption material) and environmental vibration effect are two potential challenges for EI-XPCi employing a detector with micron-scale pixels. Simulation results confirm that the behavior of an EI-XPCi system employing a high-resolution detector is susceptible to its absorption masks thickness and misalignment. This work demonstrates the potential and feasibility of employing a high-resolution direct conversion detector for phase-contrast imaging applications where higher dose efficiency, higher contrast images, and a more compact imaging system are of interest.