Evaluation of computed tomography dose profiler probe for computed tomography dose index and geometric efficiency
Philomina Akhilesh1, M S Pathan1, S D Sharma1
1Radiological Physics and Advisory Division, Bhabha Atomic Research Centre, Mumbai 400085, India.
Abstract:
The purpose of this study was to evaluate the performance of solid-state sensor based computed tomography dose profiler (CTDP) probe for measurement of standard computed tomography dose metric CTDI100and free in air geometric efficiency for various beam widths available in a 128-slice CT scanner and also to estimate the efficiency of CTDI100metric. The response accuracy of CTDP probe was verified using a standard 100 mm long ionization chamber. The geometric efficiency measurements performed by the CTDP probe were validated using XR-QA2 radiochromic film measurements. The efficiency of CTDI100metric was assessed by calculating the ratio of CTDI100measured in the center hole position to CTDI∞measured in the same position of both head and body phantoms. The weighted CTDI values derived from CTDI100measured by the CTDP probe showed an average difference of 8% from ionization chamber measured values. The efficiency of CTDI100metric estimated using CTDP probe and 150 mm long phantoms was in the range of 82% to 86% and 76% to 80% for head and body phantom measurements respectively. The differences in the geometric efficiency values for various beam settings and tube voltages measured by the CTDP probe and films were within 7%. Taken together, the results of this study proved that unlike the 100 mm long ionization chamber, the CTDP probe can be efficiently used to determine CTDI for any length over which dose integration is desired and also measure geometric efficiency of MDCT scanners for various beam widths in helical mode of operation.
More Related Videos
06:20Irradiator Commissioning and Dosimetry for Assessment of LQ α and β Parameters, Radiation Dosing Schema, and in vivo Dose Deposition
Published on: March 11, 2021
10:33Expedited Radiation Biodosimetry by Automated Dicentric Chromosome Identification ADCI and Dose Estimation
Published on: September 4, 2017
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies III: Computed Tomography