Perimetric Comparison Between the IMOvifa and Humphrey Field Analyzer
Takashi Nishida1, Medi Eslani, Robert N Weinreb
1Viterbi Family Department of Ophthalmology, Hamilton Glaucoma Center, Shiley Eye Institute, University of California San Diego, La Jolla, CA.
The IMO visual function analyzer (IMOvifa) significantly reduces measurement time for visual field testing compared to the Humphrey Field Analyzer (HFA). Results show good agreement for key metrics, suggesting a potential for reduced patient and examiner fatigue.
Area of Science:
- Ophthalmology
- Medical Technology
Background:
- The Humphrey Field Analyzer (HFA) is a standard for automated perimetry.
- Faster visual field (VF) testing methods are needed to improve patient comfort and clinic efficiency.
Purpose of the Study:
- To evaluate the performance of the IMO visual function analyzer (IMOvifa) for binocular visual field testing.
- To compare IMOvifa's results against standard automated perimetry (HFA).
Main Methods:
- A cross-sectional study evaluated 138 eyes using both HFA 24-2 SITA-Fast and IMOvifa 24-2 AIZE-Rapid.
- Key metrics including Mean Deviation (MD), Pattern Standard Deviation (PSD), foveal threshold, and Visual Field Index (VFI) were compared.
- Measurement time was recorded for each device.
Main Results:
- IMOvifa testing was significantly faster (256s vs. 419s, P <0.001), reducing measurement time by 39%.
- No significant differences were found in MD and VFI between the devices (P >0.05).
- Strong correlations were observed for MD (r=0.90), PSD (r=0.78), and VFI (r=0.94).
Conclusions:
- IMOvifa offers a faster alternative to HFA for visual field testing with comparable results for MD and VFI.
- The reduced testing time may decrease patient and examiner fatigue.
- Further research is recommended to assess IMOvifa's utility in routine clinical practice.