Related Experiment Video
Updated: Aug 10, 2025

High-resolution Functional Magnetic Resonance Imaging Methods for Human Midbrain
Published on: May 10, 2012
Improving detection of fMRI activation at 1.5 T using high permittivity ceramics
Vladislav Koloskov1, Mikhail Zubkov1, Georgiy Solomakha1
1School of Physics and Engineering, ITMO University, St. Petersburg, Russian Federation.
Abstract:
In this work, we propose an application of high permittivity materials (HPMs) to improve functional magnetic resonance imaging (fMRI) at 1.5 T, increasing the receive (Rx) sensitivity of a commercial multi-channel head coil. To evaluate the transmit efficiency, specific absorption rate (SAR), and the signal-to-noise ratio (SNR) changes introduced by the HPMs with relative permittivity of 4500, we considered the following configurations in simulation: a whole-body birdcage coil and an Rx-only multi-channel head coil with and without the HPM blocks in the presence of a homogeneous head phantom or a human body model. Experimental studies were also performed with a phantom and with volunteers. Seven healthy volunteers enrolled in a prospective study of fMRI activation in the motor cortex with and without HPMs. fMRI data were analyzed using group-level paired T-tests between acquisitions with and without HPM blocks. Both electromagnetic simulations and experimental measurements showed ∼25% improvement in the Rx sensitivity of a commercial head coil in the areas of interest when HPM blocks were placed in close proximity. It increased the detected motor cortex fMRI activation volume by an average of 56%, thus resulting in more sensitive functional imaging at 1.5 T.