Related Experiment Video
Updated: Aug 1, 2025

MRI Mapping of Cerebrovascular Reactivity via Gas Inhalation Challenges
Published on: December 17, 2014
Ecodesign and Operational Strategies to Reduce the Carbon Footprint of MRI for Energy Cost Savings
Sean A Woolen1, Amy E Becker1, Alastair J Martin1
1From the Department of Radiology and Biomedical Imaging, University of California-San Francisco, 505 Parnassus Ave, San Francisco, CA 94143 (S.A.W., A.E.B., A.J.M., C.P.H.); Siemens Healthineers, Erlangen, Germany (R.K.); Siemens Smart Infrastructure, Fremont, Calif (V.L.); Siemens Smart Infrastructure, Austin, Tex (J.F.); and Siemens Healthineers, Malvern, Pa (C.E., V.D.).
Abstract:
Background Radiology is a major contributor to health care's climate footprint due to energy-intensive devices, particularly MRI, which uses the most energy. Purpose To determine the energy, cost, and carbon savings that could be achieved through different scanner power management strategies. Materials and Methods In this retrospective evaluation, four outpatient MRI scanners from three vendors were individually equipped with power meters (1-Hz sampling rate). Power measurement logs were extracted for 39 days. Data were segmented into off, idle, prepared-to-scan, scan, or power-save modes for each scanner. Energy, cost (assuming a mean cost of $0.14 per kilowatt hour), and carbon savings were calculated for the lowest scanner activity modes. Data were summarized using descriptive statistics and 95% CIs. Results Projected annual energy consumption per scanner ranged from 82.7 to 171.1 MW-hours, with 72%-91% defined as nonproductive. Power draws for each mode were measured as 6.4 kW ± 0.1 (SD; power-save mode), 7.3 kW ± 0.6 to 9.7 kW ± 0.2 (off), 9.5 kW ± 0.9 to 14.5 kW ± 0.5 (idle), 17.3 kW ± 0.5 to 25.6 kW ± 0.6 (prepared-to-scan mode), and 28.6 kW ± 8.6 to 48.3 kW ± 11.8 (scan mode). Switching MRI units from idle to off mode for 12 hours overnight reduced power consumption by 25%-33%, translating to a potential annual savings of 12.3-21.0 MW-hours, $1717-$2943, and 8.7-14.9 metric tons of carbon dioxide (CO2) equivalent (MTCO2eq). The power-save mode further reduced consumption by 22%-28% compared with off mode, potentially saving an additional 8.8-11.4 MW-hours, $1226-$1594, and 6.2-8.1 MTCO2eq per year for 12 hours overnight. Implementation of a power-save mode for 12 hours overnight in all outpatient MRI units in the United States could save U.S. health care 58 863.2-76 288.2 MW-hours, $8.2-$10.7 million, and 41 606.4-54 088.3 MTCO2eq. Conclusion Powering down MRI units made radiology departments more energy efficient and showed substantial sustainability and cost benefits. © RSNA, 2023 Supplemental material is available for this article. See also the article by Vosshenrich and Heye in this issue.
More Related Videos
08:58Use of a Foot-Induced Digitally Controlled Resistance Device for Functional Magnetic Resonance Imaging Evaluation in Patients with Foot Paresis
Published on: July 7, 2023
08:35Oxygenation-sensitive Cardiac MRI with Vasoactive Breathing Maneuvers for the Non-invasive Assessment of Coronary Microvascular Dysfunction
Published on: August 17, 2022
Related Concept Videos
Magnetic Resonance Imaging
Imaging Studies for Cardiovascular System IV: CMRI
Radiological Investigation II: MRI and Ventilation Perfusion Scan
Magnetic Resonance Imaging (MRI) and Ventilation Perfusion Scans are two radiological investigations that offer detailed diagnostic images of the body, particularly lung structures.
MRI
MRI uses magnetic fields and radiofrequency signals to distinguish between normal and abnormal tissues. This technology provides a more detailed diagnostic image than CT scans, enabling it to characterize pulmonary nodules, stage bronchogenic carcinoma, and evaluate inflammatory activity in...
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Imaging Studies IV: Magnetic Resonance Imaging