Integrated membrane-free thermal flow sensor for silicon-on-glass microfluidics
Vitaly V Ryzhkov1, Vladimir V Echeistov1,2, Aleksandr V Zverev1
1FMN Laboratory, Bauman Moscow State Technical University, Moscow, 105005, Russia. irodionov@bmstu.ru.
Lab on a Chip
|May 18, 2023
Summary
A novel membrane-free microfluidic thermal flow sensor (MTFS) is integrated directly into silicon-glass lab-on-a-chip devices. This sensor offers precise, real-time liquid flow control for portable analytical systems, overcoming limitations of external flow meters.
Area of Science:
- Microfluidics
- Sensor Technology
- Analytical Instrumentation
Background:
- Lab-on-a-chip (LOC) systems require precise control of ultralow liquid reagent flows for multistep reactions.
- Existing flow meters often introduce dead volume and are incompatible with microfluidic chip fabrication.
- A need exists for integrated, robust flow sensors for advanced portable analytical devices.
Purpose of the Study:
- To develop and characterize a novel membrane-free microfluidic thermal flow sensor (MTFS) for seamless integration into LOC devices.
- To propose design rules for optimizing MTFS sensitivity and measurement range.
- To demonstrate the compatibility of the MTFS with corrosive liquids and its long-term stability.
Main Methods:
- Fabrication of a membrane-free MTFS using a 4'' wafer silicon-glass process, integrating thin-film thermo-resistive elements isolated from microchannels.
- Development of design rules for maximizing sensor performance.
- Implementation of an automated calibration method for the thermo-resistive elements.
- Experimental validation against a Coriolis flow sensor over extended periods.
Main Results:
- Successful integration of the MTFS into silicon-glass microfluidic chips.
- Demonstrated compatibility with corrosive liquids, crucial for biological applications.
- Achieved a relative flow error of less than 5% for flow rates between 2-30 μL min⁻¹.
- Exhibited a sub-second time response and stability over hundreds of hours of testing.
Conclusions:
- The developed membrane-free MTFS is a viable, integrated solution for precise liquid flow control in LOC devices.
- This technology enhances the capabilities of portable analytical systems, particularly for applications involving corrosive reagents.
- The MTFS offers a robust, accurate, and fast alternative to conventional external flow meters in microfluidic applications.