Related Experiment Video
Updated: Aug 3, 2026

An Alternative to the Traditional Cold Pressor Test: The Cold Pressor Arm Wrap
Published on: January 16, 2014
The Certification Matters: A Comparative Performance Analysis of Combat Application Tourniquets versus Non-Certified
Emanuele Lagazzi1,2,3, Giulia Ballardini4,5, Alberto Drogo3,6
1Department of Surgery, Humanitas Research Hospital, Rozzano, Lombardia, Italy.
Introduction:
The Stop the Bleed campaign provided civilians with tourniquet application training and increased the demand for tourniquets among the general population, which led to the development of new commercially available devices. However, most widely available tourniquets have not undergone testing by regulatory bodies and their efficacy remains unknown.
Study Objective:
This study aimed to compare the efficacy and performance of Combat Application Tourniquets (CAT) versus uncertified tourniquets.
Methods:
This study compared 25 CAT with 50 commercially available "look-alike" tourniquets (LA-TQ) resembling the CAT. The CAT and the LA-TQ were compared for cost, size, and tested during one-hour and six-hour applications on a manikin's leg. The outcomes were force applied, force variation during the application, and tourniquet rupture rate.
Results:
The LA-TQ were cheaper (US$6.07 versus US$27.19), shorter, and had higher inter-device variability than the CAT (90.1 [SE = 0.5] cm versus 94.5 [SE = 0.1] cm; P <.001). The CAT applied a significantly greater force during the initial application when compared to the LA-TQ (65 [SE = 3] N versus 14 [SE = 1] N; P <.001). While the initial application force was maintained for up to six hours in both groups, the CAT group applied an increased force during one-hour applications (group effect: F [1,73] = 105.65; P <.001) and during six-hour applications (group effect: F [1,12] = 9.79; P = .009). The rupture rate differed between the CAT and the LA-TQ (0% versus 4%).
Conclusion:
The LA-TQ applied a significantly lower force and had a higher rupture rate compared to the CAT, potentially affecting tourniquet performance in the context of public bleeding control. These findings warrant increased layperson education within the framework of the Stop the Bleed campaign and further investigations on the effectiveness of uncertified devices in real-world applications.
Related Concept Videos
Multiple Comparison Tests
It would be easy to compare two samples using a significance alpha level of 0.05. In other words, there is only one sample pair to be compared. However, it would be difficult to identify a significantly different sample if the number...
Equivalent Couples
Two couples are considered to be equivalent if they produce the same rotational effect on a rigid body. In other words, the two couples have the same magnitude and act in the same direction, causing the same angular displacement or acceleration in the body.
For instance, consider two couples lying in the plane of the page, with one having a pair of equal...
Relative Motion Analysis using Rotating Axes-Problem Solving
Here, in order to determine the magnitude of velocity and acceleration for point...