Image quality improvement in bowtie-filter-equipped cone-beam CT using a dual-domain neural network
Sungho Yun1, Uijin Jeong1, Donghyeon Lee1
1Department of Nuclear and Quantum Engineering, Korea Advanced Institute of Science and Technology (KAIST), Daejeon, South Korea.
Background:
The bowtie-filter in cone-beam CT (CBCT) causes spatially nonuniform x-ray beam often leading to eclipse artifacts in the reconstructed image. The artifacts are further confounded by the patient scatter, which is therefore patient-dependent as well as system-specific.
Purpose:
In this study, we propose a dual-domain network for reducing the bowtie-filter-induced artifacts in CBCT images.
Methods:
In the projection domain, the network compensates for the filter-induced beam-hardening that are highly related to the eclipse artifacts. The output of the projection-domain network was used for image reconstruction and the reconstructed images were fed into the image-domain network. In the image domain, the network further reduces the remaining cupping artifacts that are associated with the scatter. A single image-domain-only network was also implemented for comparison.
Results:
The proposed approach successfully enhanced soft-tissue contrast with much-reduced image artifacts. In the numerical study, the proposed method decreased perceptual loss and root-mean-square-error (RMSE) of the images by 84.5% and 84.9%, respectively, and increased the structure similarity index measure (SSIM) by 0.26 compared to the original input images on average. In the experimental study, the proposed method decreased perceptual loss and RMSE of the images by 87.2% and 92.1%, respectively, and increased SSIM by 0.58 compared to the original input images on average.
Conclusions:
We have proposed a deep-learning-based dual-domain framework to reduce the bowtie-filter artifacts and to increase the soft-tissue contrast in CBCT images. The performance of the proposed method has been successfully demonstrated in both numerical and experimental studies.
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies III: Computed Tomography