Oversensing caused by trapped air in the header of subcutaneous cardioverter-defibrillator generator
Takuya Okada1, Junji Morita2, Takayuki Kitai2
1Department of Clinical Engineering, Sapporo Heart Center Sapporo Cardiovascular Clinic Sapporo Japan.
Abstract:
In this study, we report two cases with oversensing due to air accumulation in the subcutaneous implantable cardioverter-defibrillator (S-ICD) device generator header. If trapped air in the header of the device is suspected, the re-connection procedure should be considered or the primary vector must be used, which might prevent oversensing.
More Related Videos
12:45Benefits of Cardiac Resynchronization Therapy in an Asynchronous Heart Failure Model Induced by Left Bundle Branch Ablation and Rapid Pacing
Published on: December 11, 2017
07:39Use of a Percutaneous Ventricular Assist Device/Left Atrium to Femoral Artery Bypass System for Cardiogenic Shock
Published on: August 16, 2021
Related Concept Videos
Generator Voltage Control
Imbalances in Cardiac Output
CHF can occur due to the failure of either side of the heart. Left-side failure leads to pulmonary congestion—the right side continues to send...
Cardiopulmonary Resuscitation III: AED Use
Van de Graaff Generator
Van de Graaff uses both smooth and pointed surfaces, conductors, and insulators to generate large static charges and, hence, large voltages. A substantial excess charge can be deposited on the sphere because it moves...
Cardiomyopathy II: Dilated Cardiomyopathy
DC Generator