Scattered Radiation Distribution Utilizing Three Different Cone-Beam Computed Tomography Devices for Maxillofacial
Sotirios Petsaros1, Emmanouil Chatzipetros1, Catherine Donta1
1Department of Oral Diagnosis and Radiology, Faculty of Dentistry, National and Kapodistrian University of Athens, 2 Thivon Street, Goudi, 11527 Athens, Greece.
Abstract:
This study aimed to estimate scattered radiation and its spatial distribution around three cone-beam computed tomography (CBCT) devices, in order to determine potential positions for an operator to stand if they needed to be inside the CBCT room. The following devices were tested: Morita Accuitomo (CBCT1), Newtom Giano HR (CBCT2), Newtom VGi (CBCT3). Scattered radiation measurements were performed using different kVp, mA, and Field of View (FOV) options. An anthropomorphic phantom (NATHANIA) was placed inside the X-ray gantry to simulate clinical conditions. Scattered measurements were taken with the Inovision model 451P Victoreen ionization chamber once placed at fixed distances from each irradiation isocenter, away from the primary beam. A statistically significant (p < 0.001) difference was found in the mean value of the scattered radiation estimations between the CBCT devices. Scattered radiation was reduced with a different rate for each CBCT device as distance was increased. For CBCT1 the reduction was 0.047 μGy, for CBCT2 it was 0.036 μGy, and for CBCT3 it was 0.079 μGy, for every one meter from the X-ray gantry. Therefore, at certain distances from the central X-ray, the scattered radiation was below the critical level of 1 mGy, which is defined by the radiation protection guidelines as the exposure radiation limit of the general population. Consequently, an operator could stay inside the room accompanying the patient being scanned, if necessary.
More Related Videos
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies III: Computed Tomography
Radiological Investigation I: X-ray and CT
X-ray Imaging
Radiological Investigation II: MRI and Ventilation Perfusion Scan
Magnetic Resonance Imaging (MRI) and Ventilation Perfusion Scans are two radiological investigations that offer detailed diagnostic images of the body, particularly lung structures.
MRI
MRI uses magnetic fields and radiofrequency signals to distinguish between normal and abnormal tissues. This technology provides a more detailed diagnostic image than CT scans, enabling it to characterize pulmonary nodules, stage bronchogenic carcinoma, and evaluate inflammatory activity in...
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...