Evaluating In Vitro Performance of a Novel Stainless Steel Rotary System (Gentlefile) Based on Debris Extrusion and
Ahmad Nouroloyouni1, Fatemeh Safavi Hir1, Robab Farhang1
1Department of Endodontics, School of Dentistry, Ardabil University of Medical Science, Ardabil, Iran.
Biomed Research International
|November 8, 2023
Summary
The Gentlefile (GF) system did not reduce debris extrusion or instrumentation time compared to ProTaper Universal (PTU) nickel-titanium files. Manual step-back (MSB) techniques took significantly longer for root canal instrumentation.
Area of Science:
- Endodontics
- Dental Materials Science
Background:
- Nickel-titanium (NiTi) files offer mechanical advantages in endodontic procedures.
- The new Gentlefile (GF) stainless steel system claims superiority over traditional NiTi files.
- Comparative studies are essential to validate new endodontic instrument claims.
Purpose of the Study:
- To compare the efficacy of the Gentlefile (GF) system against the ProTaper Universal (PTU) NiTi system and a manual step-back (MSB) stainless steel technique.
- To evaluate debris extrusion and instrumentation time for each root canal instrumentation method.
- To assess the performance of the GF system in relation to established endodontic techniques.
Main Methods:
- An in vitro study using 66 human mandibular premolars with minimal root curvature.
- Teeth were randomly assigned to three groups (n=22): PTU, GF, or MSB instrumentation.
- Apical debris extrusion was quantified by weighing pre- and post-instrumentation vials; instrumentation time was recorded.
Main Results:
- No statistically significant difference in apical debris extrusion was observed among the PTU, GF, and MSB groups (P > 0.05).
- A significant difference in instrumentation time was found between the groups (P < 0.05).
- MSB technique required significantly longer instrumentation time compared to both PTU and GF systems (P < 0.05).
Conclusions:
- The Gentlefile (GF) system did not demonstrate superior performance in reducing apical debris extrusion or instrumentation time compared to the ProTaper Universal (PTU) system.
- The manual step-back (MSB) technique was associated with the longest instrumentation time.
- Further research may be needed to explore potential benefits or limitations of the GF system in different clinical scenarios.