A Bioimpedance Spectroscopy Interface for EIM Based on IF-Sampling and Pseudo 2-Path SC Bandpass ∆Σ ADC
IEEE Transactions on Biomedical Circuits and Systems
|February 26, 2024
Summary
This study introduces a low-noise bioimpedance spectroscopy interface for electrical impedance myography. The system accurately detects muscle contractions, showing potential for wearable EIM devices.
Area of Science:
- Biomedical Engineering
- Electrical Engineering
- Signal Processing
Background:
- Electrical Impedance Myography (EIM) requires accurate bioimpedance measurements.
- Existing interfaces face challenges in noise, power efficiency, and data throughput for detecting muscle contractions.
Purpose of the Study:
- To present a novel low-noise bioimpedance spectroscopy interface for EIM.
- To achieve high accuracy and efficiency in bioimpedance readings across a wide frequency range.
- To demonstrate the interface's capability for detecting muscle activity.
Main Methods:
- Utilized a direct-digital-synthesis (DDS) sinusoidal signal generator for accurate bioimpedance readings.
- Employed a quadrature low-intermediate frequency (IF) readout with low-gain IAs and bandpass ∆Σ ADCs.
- Implemented the interface in a 180 nm CMOS process.
Main Results:
- Achieved 0.7 mΩ/Hz sensitivity at 15.625 kHz and 105.8 dB SNR within a 4 Hz bandwidth.
- Demonstrated a figure-of-merit of 146.5 dB, indicating excellent performance.
- Successfully recorded EIM data during bicep brachii muscle contractions.
Conclusions:
- The proposed bioimpedance spectroscopy interface offers a low-noise, efficient solution for EIM.
- The system's performance is suitable for detecting muscle contractions in real-time.
- This technology holds significant potential for developing advanced wearable EIM systems.
Related Concept Videos
Bandpass Sampling
180
In signal processing, bandpass sampling is an effective technique for sampling signals that have most of their energy concentrated within a narrow frequency band. This type of signal is known as a bandpass signal. The key principle of bandpass sampling involves sampling the signal at a rate that is greater than twice the signal's bandwidth to prevent aliasing.
A bandpass signal has a spectrum with a lower frequency limit, denoted as ω1, and an upper frequency limit, denoted as ω2....
A bandpass signal has a spectrum with a lower frequency limit, denoted as ω1, and an upper frequency limit, denoted as ω2....
180
Instrumentation Amplifier
517
An electrocardiography (ECG) machine is an essential piece of medical equipment used to monitor the electrical activity of the heart. It operates by detecting small electrical changes on the skin that result from the depolarization of the heart muscle during each heartbeat. However, these signals are in the microvolt range and can be easily overwhelmed by noise or interference.
To overcome this challenge, an ECG machine utilizes an instrumentation amplifier. This specialized amplifier is...
To overcome this challenge, an ECG machine utilizes an instrumentation amplifier. This specialized amplifier is...
517
Bode Plots Construction
706
The Bode plot is an essential tool in control system analysis, mapping the frequency response of a system through a magnitude plot and a phase plot, both against a logarithmic frequency axis. To construct a Bode plot, consider the transfer function H(ω):
706
Inductively Coupled Plasma Atomic Emission Spectroscopy: Instrumentation
221
Inductively coupled plasma (ICP) is the common plasma source used in atomic emission spectroscopy (AES), a technique that detects and analyzes various elements in a sample. This method is often called inductively coupled plasma atomic emission spectroscopy (ICP-AES).
There are three main types of inductively coupled plasma atomic emission spectroscopy (ICP-AES) instruments: sequential, simultaneous multichannel, and Fourier transform instruments, with the latter being less commonly used....
There are three main types of inductively coupled plasma atomic emission spectroscopy (ICP-AES) instruments: sequential, simultaneous multichannel, and Fourier transform instruments, with the latter being less commonly used....
221