[Validation of Optimal Imaging Conditions for Coronary Computed Tomography Angiography Using High-definition Mode and
Nobuo Kitera1, Chikako Fujioka1, Toru Higaki2
1Department of Radiology, Hiroshima University Hospital.
Purpose:
To verify the optimal imaging conditions for coronary computed tomography angiography (CCTA) examinations when using high-definition (HD) mode and deep learning image reconstruction (DLIR) in combination.
Method:
A chest phantom and an in-house phantom using 3D printer were scanned with a 256-row detector CT scanner. The scan parameters were as follows - acquisition mode: ON (HD mode) and OFF (normal resolution [NR] mode), rotation time: 0.28 s/rotation, beam coverage width: 160 mm, and the radiation dose was adjusted based on CT-AEC. Image reconstruction was performed using ASiR-V (Hybrid-IR), TrueFidelity Image (DLIR), and HD-Standard (HD mode) and Standard (NR mode) reconstruction kernels. The task-based transfer function (TTF) and noise power spectrum (NPS) were measured for image evaluation, and the detectability index (d') was calculated. Visual evaluation was also performed on an in-house coronary phantom.
Result:
The in-plane TTF was better for the HD mode than for the NR mode, while the z-axis TTF was lower for DLIR than for Hybrid-IR. The NPS values in the high-frequency region were higher for the HD mode compared to those for the NR mode, and the NPS was lower for DLIR than for Hybrid-IR. The combination of HD mode and DLIR showed the best value for in-plane d', whereas the combination of NR mode and DLIR showed the best value for z-axis d'. In the visual evaluation, the combination of NR mode and DLIR showed the best values from a noise index of 45 HU.
Conclusion:
The optimal combination of HD mode and DLIR depends on the image noise level, and the combination of NR mode and DLIR was the best imaging condition under noisy conditions.
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...
Imaging Studies for Cardiovascular System IV: CMRI
Imaging Studies for Cardiovascular System V: CT
Imaging Studies for Cardiovascular System VI: Calcium -Scoring CT
Imaging Studies III: Computed Tomography