Related Experiment Video
Updated: Jun 29, 2025

Author Spotlight: Advancements in Impedance Monitoring for Cochlear Implant Surgery
Published on: August 4, 2023
Towards inferring positioning of straight cochlear-implant electrode arrays during insertion using real-time
Katherine E Riojas1, Trevor L Bruns1, Josephine Granna1
1Department of Mechanical Engineering, Vanderbilt University, Nashville, Tennessee, USA.
Background:
Cochlear-implant electrode arrays (EAs) are currently inserted with limited feedback, and impedance sensing has recently shown promise for EA localisation.
Methods:
We investigate the use of impedance sensing to infer the progression of an EA during insertion.
Results:
We show that the access resistance component of bipolar impedance sensing can detect when a straight EA reaches key anatomical locations in a plastic cochlea and when each electrode contact enters/exits the cochlea. We also demonstrate that dual-sided electrode contacts can provide useful proximity information and show the real-time relationship between impedance and wall proximity in a cadaveric cochlea for the first time.
Conclusion:
The access resistance component of bipolar impedance sensing has high potential for estimating positioning information of EAs relative to anatomy during insertion. Main limitations of this work include using saline as a surrogate for human perilymph in ex vivo models and using only one type of EA.