Related Experiment Video
Updated: Jun 28, 2025

Frequency Mixing Magnetic Detection Scanner for Imaging Magnetic Particles in Planar Samples
Published on: June 9, 2016
Segmented RF shield design to minimize eddy currents for low-field Halbach MRI systems
Bart de Vos1, Rob Remis2, Andrew Webb1
1C.J. Gorter MRI center, Radiology, Leiden University Center, Leiden, The Netherlands; Terahertz Sensing, Microelectronics, Delft University of Technology, Delft, The Netherlands.
A novel segmented radiofrequency (RF) shield for low-field MRI significantly reduces eddy current decay times by up to 2.9x. This improved shield minimizes image blurring in MRI scans without increasing noise interference.
Area of Science:
- Medical Imaging
- Electrical Engineering
- Biophysics
Background:
- Radiofrequency (RF) coils in MRI systems require shielding to minimize noise and coil coupling.
- Existing shields can impede gradient fields, leading to undesirable eddy currents.
- Low-field, point-of-care MRI devices present unique design challenges for effective shielding.
Purpose of the Study:
- To design and evaluate a novel segmented double-layer RF shield for low-field Halbach-based MRI systems.
- To minimize eddy currents while maintaining RF shielding effectiveness.
- To assess the impact of the new shield on image quality.
Main Methods:
- Eddy current simulations were performed to guide the shield design.
- A segmented double-layer shield was constructed based on simulation results.
- Shield performance was evaluated by measuring eddy current decay times and RF noise coupling.
- Turbo spin echo MRI scans of a phantom and a healthy volunteer brain were acquired.
Main Results:
- The segmented shield achieved a maximum reduction factor of 2.9 in eddy current decay time compared to a reference shield.
- The segmented shield exhibited comparable RF noise coupling to the unsegmented reference shield.
- MRI images demonstrated reduced blurring when using the segmented shield.
Conclusions:
- The developed segmented double-layer shield effectively reduces eddy currents in low-field MRI.
- This design offers improved image quality by minimizing blurring without compromising noise performance.
- The shield is suitable for low-field, point-of-care Halbach-based MRI applications.
Related Concept Videos
Diamagnetic Shielding of Nuclei: Local Diamagnetic Current
Eddy Currents
Other major applications of eddy currents appear in metal detectors and the braking systems of trains and roller...
Magnetic Field Of A Current Loop
Magnetic Resonance Imaging
Magnetic Damping
If, however, the bob is a slotted metal plate, the magnet produces a much smaller effect. When a slotted metal plate enters the field, an emf is induced by the change in flux; however, it is less effective because the slots limit the...
NMR Spectrometers: Resolution and Error Correction

