Jove
Visualize
Contact Us
JoVE
x logofacebook logolinkedin logoyoutube logo
ABOUT JoVE
OverviewLeadershipBlogJoVE Help Center
AUTHORS
Publishing ProcessEditorial BoardScope & PoliciesPeer ReviewFAQSubmit
LIBRARIANS
TestimonialsSubscriptionsAccessResourcesLibrary Advisory BoardFAQ
RESEARCH
JoVE JournalMethods CollectionsJoVE Encyclopedia of ExperimentsArchive
EDUCATION
JoVE CoreJoVE BusinessJoVE Science EducationJoVE Lab ManualFaculty Resource CenterFaculty Site
Terms & Conditions of Use
Privacy Policy
Policies

Related Concept Videos

Bode Plots Construction01:24

Bode Plots Construction

698
The Bode plot is an essential tool in control system analysis, mapping the frequency response of a system through a magnitude plot and a phase plot, both against a logarithmic frequency axis. To construct a Bode plot, consider the transfer function H(ω):
698
NMR Spectrometers: Resolution and Error Correction01:14

NMR Spectrometers: Resolution and Error Correction

686
When magnetic nuclei in a sample achieve resonance and undergo relaxation, the signal detected in NMR is an approximately exponential free induction decay. Fourier transform of an exponential decay yields a Lorentzian peak in the frequency domain. Lorentzian peaks in an NMR spectrum are defined by their amplitude, full width at half maximum, and position, where the peak width is governed by the spin-spin relaxation time alone. In real experiments, however, the applied magnetic field is rendered...
686

You might also read

Related Articles

Articles linked to this work by shared authors, journal, and citation graph.

Sort by
Same author

Optimisation of magnetic field sensing with optically pumped magnetometers for magnetic detection electrical impedance tomography.

Physics in medicine and biology·2025
Same author

Towards spatially selective efferent neuromodulation: anatomical and functional organization of cardiac fibres in the porcine cervical vagus nerve.

The Journal of physiology·2024
Same author

Anatomical and functional organization of cardiac fibers in the porcine cervical vagus nerve allows spatially selective efferent neuromodulation.

bioRxiv : the preprint server for biology·2024
Same author

Non-invasive imaging of neural activity with magnetic detection electrical impedance tomography (MDEIT): a modelling study.

Physiological measurement·2023
Same author

Overcoming temporal dispersion for measurement of activity-related impedance changes in unmyelinated nerves.

Journal of neural engineering·2022
Same author

Simplifying the hardware requirements for fast neural EIT of peripheral nerves.

Physiological measurement·2021
Same journal

Dissecting the integrated information of cardiovascular and cardiorespiratory systems at rest and during physiological stress.

Physiological measurement·2026
Same journal

Respiratory event type and duration modulate PPG waveforms in OSA.

Physiological measurement·2026
Same journal

Estimating changes in systolic blood pressure based on pulse wave morphology using paired segment comparison.

Physiological measurement·2026
Same journal

Small changes in hand height alter absorbance, but not pulsation, in the finger pulse plethysmograph.

Physiological measurement·2026
Same journal

A Comprehensive Inference-Time Augmentation Framework in Physiological Signals: Application to PPG-Based AF Detection.

Physiological measurement·2026
Same journal

Quantification of pendelluft in electrical impedance tomography data: opening Pandora's box? A literature review of analytical methods.

Physiological measurement·2026
See all related articles

Related Experiment Video

Updated: Jun 25, 2025

Monitoring Lung Function with Electrical Impedance Tomography in the Intensive Care Unit
05:56

Monitoring Lung Function with Electrical Impedance Tomography in the Intensive Care Unit

Published on: September 6, 2024

2.0K

Noise-based correction for electrical impedance tomography.

Kai Mason1, Florencia Maurino-Alperovich1, David Holder1

  • 1Department of Medical Physics and Biomedical Engineering, University College London, London, United Kingdom.

Physiological Measurement
|May 21, 2024
PubMed
Summary
This summary is machine-generated.

A new noise-based correction (NBC) method significantly enhances image quality in electrical impedance tomography (EIT) by reducing noise propagation. Uncorrelated noise-based correction is recommended for general use in EIT applications.

Keywords:
electrical impedance tomographyimage reconstructionnoise propagationpost-processing

More Related Videos

Enhancing Electrode Location Assessment in Cochlear Implantation via Computed Tomography Image Fusion
03:58

Enhancing Electrode Location Assessment in Cochlear Implantation via Computed Tomography Image Fusion

Published on: January 17, 2025

411
Author Spotlight: Advancements in Impedance Monitoring for Cochlear Implant Surgery
06:54

Author Spotlight: Advancements in Impedance Monitoring for Cochlear Implant Surgery

Published on: August 4, 2023

1.1K

Related Experiment Videos

Last Updated: Jun 25, 2025

Monitoring Lung Function with Electrical Impedance Tomography in the Intensive Care Unit
05:56

Monitoring Lung Function with Electrical Impedance Tomography in the Intensive Care Unit

Published on: September 6, 2024

2.0K
Enhancing Electrode Location Assessment in Cochlear Implantation via Computed Tomography Image Fusion
03:58

Enhancing Electrode Location Assessment in Cochlear Implantation via Computed Tomography Image Fusion

Published on: January 17, 2025

411
Author Spotlight: Advancements in Impedance Monitoring for Cochlear Implant Surgery
06:54

Author Spotlight: Advancements in Impedance Monitoring for Cochlear Implant Surgery

Published on: August 4, 2023

1.1K

Area of Science:

  • Biomedical Engineering
  • Medical Imaging
  • Electrical Engineering

Background:

  • Noisy measurements in impedance imaging lead to inaccurate images.
  • Existing post-processing techniques cannot effectively calculate and suppress measurement noise propagation.

Purpose of the Study:

  • Develop a noise-based correction (NBC) post-processing method for impedance tomography.
  • Evaluate NBC's effectiveness in improving electrical impedance tomography (EIT) image quality.
  • Compare the performance of correlated versus uncorrelated noise for NBC.
  • Assess NBC's applicability to in vivo data and its stability across different geometries.

Main Methods:

  • Performed in silico EIT in 2D and 3D domains (homogeneous and heterogeneous human head).
  • Validated methods using experimental EIT in saline tanks (2D circular and 3D human head-shaped).
  • Assessed image quality using weighted spatial variance (WSV) error.
  • Tested NBC on in vivo EIT data from human lung ventilation and rat brain activity.

Main Results:

  • NBC maintained or improved image quality visually across all tested conditions (in silico and experimental).
  • NBC significantly improved WSV in nearly all cases, indicating enhanced accuracy.
  • NBC improved in vivo EIT data quality, preserving reconstructed perturbation shapes.

Conclusions:

  • Noise-based correction (NBC) is an effective post-processing technique for improving EIT image quality.
  • Uncorrelated noise-based correction demonstrated superior performance and is recommended for general EIT applications.