Ultrasound backscatter coefficient for fat quantification is affected by the measurement depth
Giovanna Ferraioli1, Annalisa De Silvestri2, Gabriela Torres3
1Dipartimento di Scienze Clinico-Chirurgiche, Diagnostiche e Pediatriche, University of Pavia, Viale Brambilla 74, 27100, Pavia, Italy. giovanna.ferraioli@unipv.it.
Abdominal Radiology (New York)
|June 4, 2024
Summary
Ultrasound backscatter coefficient (BSC) in the liver decreases with depth, similar to attenuation coefficient. This depth dependence requires standardized protocols for accurate liver assessments.
Area of Science:
- Medical Imaging
- Biophysics
- Ultrasound Technology
Background:
- Ultrasound attenuation coefficient (AC) estimation is known to decrease with measurement depth.
- The behavior of backscatter coefficient (BSC) with depth in ultrasound imaging is not well understood.
Purpose of the Study:
- To investigate whether the backscatter coefficient (BSC) in liver ultrasound exhibits a similar depth-dependent behavior as the attenuation coefficient (AC).
- To identify factors influencing BSC estimation in liver ultrasound.
Main Methods:
- Retrospective analysis of ultrasound data from 104 individuals using a Sequoia US system with an ultrasound-derived fat fraction (UDFF) algorithm.
- BSC data were extracted, and fractional polynomial regression was employed to model the relationship between BSC and depth, considering covariates like age, sex, skin-to-liver distance, and liver stiffness.
- Regions of interest were positioned at varying depths (1.5 to 5 cm) below the liver capsule.
Main Results:
- The best-fitting model for BSC incorporated depth as a linear and square root factor.
- BSC demonstrated a significant decrease with increasing depth, with a reduction of -13.98 dB/cm-steradian for each logarithmic increase in depth (p=0.001).
- Skin-to-liver capsule distance and liver stiffness were identified as independent predictors of BSC.
Conclusions:
- Liver ultrasound backscatter coefficient (BSC) estimation is significantly influenced by measurement depth.
- A standardized ultrasound acquisition protocol is essential for reliable comparison of BSC results and monitoring changes over time.
Related Concept Videos
Ultrasonography
4.4K
Ultrasonography is an imaging technique that uses high-frequency sound waves to visualize the body's internal structures. It is a non-invasive and safe procedure that does not involve the use of ionizing radiation, making it widely used in various medical fields. Ultrasonography is used to study heart function, blood flow in the neck or extremities, certain conditions such as gallbladder disease, and fetal growth and development.
During an ultrasonography procedure, a handheld device called...
During an ultrasonography procedure, a handheld device called...
4.4K
Ultrasound II: Endoscopic Ultrasound and FibroScan
97
Endoscopic Ultrasound (EUS) and FibroScan are valuable diagnostic tools in gastroenterology and hepatology, each with specific applications and techniques.
Endoscopic Ultrasound (EUS):
Endoscopic Ultrasound (EUS):
97