New type of filaments for improved accuracy of multiple sulfur isotope analyses by electron-impact gas-source mass spectrometry
Related Experiment Video
Updated: Jun 14, 2025

Preparation of Authigenic Pyrite from Methane-bearing Sediments for In Situ Sulfur Isotope Analysis Using SIMS
Published on: August 31, 2017
Pierre Cartigny1, Amaury Bouyon1,2, Hartmut Bars3
1Université de Paris, Institut de Physique du Globe de Paris, CNRS, Paris, France.
Rationale:
The analysis of the three sulfur stable isotope ratios (33S/32S, 34S/32S, 36S/32S) is routinely performed by gas-source isotope ratio mass spectrometry (IRMS) on the SF6 gaseous molecule, collecting SF5 + ions at m/z ~ 127, 128, 129 and 131. High precision and accuracy are commonly achieved owing to a lack of correction because fluorine has only one isotope and the inert nature of the SF6 molecule. The analysis of the 36S/32S ratio is, however, complicated by the low abundance of 36S (~0.015%) and the possible occurrence of trace amounts of fluorocarbon compounds leading to 12C3F5 + ions at m/z ~ 131, i.e. where 36SF5 + ions are collected.
Methods:
We used gas-source high-resolution IRMS to better characterize the nature of possible interferences, and we tested novel types of filaments in order to investigate their influence on possible interferences.
Results:
We confirm that the 12C3F5 + ion represents the main isobaric interference at m/z ~ 131. We also demonstrate that tungsten fluoride adducts are formed from the reaction of fluorine ions derived during fragmentation of the SF6 molecule with the hot tungsten filament. These reactions lead to the formation of e.g. WF5 +, WF4 +, WF3 + ions, including doubly charged ions. WF4 ++, in particular, leads to isobaric interference on m/z ~ 128, 129 and 131 from 180WF4 ++, 182WF4 ++ and 186 WF4 ++ ions, respectively. Because 180W (0.12%) is at low abundance, its influence on δ33S measurements would remain negligible, but 182W (26.5%) and 186W (28.4%) lead to scale contraction for both δ34S and δ36S.
Conclusions:
Rather than correcting for these interferences, or working at high mass resolution, we suggest avoiding W isobaric interferences by using other types of filaments, with initial reports on both pure Re filaments and Y2O3-coated W filaments.
More Related Videos
05:48Analysis of Volatile and Oxidation Sensitive Compounds Using a Cold Inlet System and Electron Impact Mass Spectrometry
Published on: September 5, 2014
09:34Improved Polymerase Chain Reaction-restriction Fragment Length Polymorphism Genotyping of Toxic Pufferfish by Liquid Chromatography/Mass Spectrometry
Published on: September 20, 2016
Related Concept Videos
Electrospray Ionization (ESI) Mass Spectrometry
ESI utilizes electrical energy to transfer ions from the liquid phase of the sample into the...
Mass Spectrometers
Mass Spectrometry: Isotope Effect
Mass Spectrometry: Overview
Atomic Emission Spectroscopy: Instrumentation
Atomic Emission Spectroscopy: Lab