A New Single Chamber Implantable Defibrillator with Atrial Sensing: A Practical Demonstration of Sensing and Ease of Implantation
Published on: February 28, 2012
High predicted durability for the novel small-diameter OmniaSecure defibrillation lead
George H Crossley1, Pamela K Mason2, Bert Hansky3
1Vanderbilt University Medical Center, Nashville, Tennessee.
The novel OmniaSecure defibrillation lead shows promising long-term durability. Reliability modeling projects a 98.2% fracture-free survival rate at 10 years, offering a reliable option for patients needing implantable cardioverter-defibrillators.
Area of Science:
- Cardiovascular medicine
- Biomedical engineering
- Medical device technology
Background:
- Implantable cardioverter-defibrillator (ICD) leads are a critical component, yet lead failure remains a concern.
- Increasing patient longevity and ICD battery life necessitate more durable defibrillation leads.
- The LEADR trial introduced the OmniaSecure lead, a small-diameter, lumenless design, showing initial safety and efficacy with no fractures in early follow-up.
Purpose of the Study:
- To project the 10-year fracture-free survival rate of the OmniaSecure defibrillation lead.
- To utilize advanced reliability modeling to predict long-term lead performance.
- To compare projected durability against existing lead technologies.
Main Methods:
- Employed a validated reliability model integrating patient data and bench testing results.
- Assessed lead bending curvature in vivo using biplane fluoroscopy during patient and cardiac motion.
- Conducted bench tests simulating in vivo conditions with exaggerated bending to determine lead fatigue strength.
Main Results:
- Reliability modeling predicts a 98.2% fracture-free survival rate for the OmniaSecure lead at 10 years.
- The projected 10-year fracture-free survival rate in adolescents is 97.9%.
- These projections surpass the modeled and clinical 10-year performance of the larger Sprint Quattro lead.
Conclusions:
- Modeling indicates the OmniaSecure lead possesses high projected durability for 10-year performance.
- The lead's reliability is particularly promising for the active adolescent pediatric population.
- The novel 4.7F defibrillation lead is designed for enhanced reliability, addressing a clinical need.
Related Concept Videos
Batteries and Fuel Cells
Turnover Number and Catalytic Efficiency
Chymotrypsin is a pancreatic enzyme that breaks down proteins during digestion. The...
Internal Combustion Engine
Mechanical Efficiency of Real Machines
However, in reality, no machine can be truly ideal, and all of them experience some...
Design Example: Automobile Ignition System
One can generate a large voltage using a car battery of 12 volts with the help of inductors. Inductors are known for opposing rapid...
Abrasion Resistance of Concrete
One such test is the revolving disc test, where three plates...