Related Experiment Video
Updated: Jun 6, 2025

Utilizing Percutaneous Ventricular Assist Devices in Acute Myocardial Infarction Complicated by Cardiogenic Shock
Published on: June 12, 2021
Assessment of hemodynamic parameters by PiCCO and PAC in patients treated with the Impella CP
Dominik Schmitt1, Andreas Schneider1, Björn Lengenfelder1,2
1Department of Internal Medicine I, Medical Intensive Care Unit, University Hospital, University of Wuerzburg, Wuerzburg, Germany.
Abstract:
In 28 patients supported by an Impella pump (Impella CP, Abiomed Inc.®, Danvers, MA), hemodynamic measurements by PiCCO (PULSION Medical Systems SE®, Feldkirchen, Germany) and pulmonary artery catheter (PAC or Swan-Ganz Catheter; Edwards Lifescience®, Unterschleissheim, Germany) were compared. There was a significant positive correlation of cardiac output (CO; r 2 = 0.917, p < 0.001), systemic vascular resistance index (SVRI; r 2 = 0.904, p < 0.001), stroke volume index (SI; r 2 = 0.909, p < 0.001) and left ventricular work index (LCWI; r 2 = 0.689, p < 0.001) in PiCCO and PAC measurements under Impella CP support. We conclude, that in patients with left-ventricular Impella support, hemodynamics may be assessentd by PiCCO or PAC.