[Determination of trace anions in G2-grade phosphoric acid using two-dimensional ion-exclusion-ion-exchange
Jia-Ding Wang1,2, Gao Tang1, Yan Lu2
1College of Materials and Chemistry, China Jiliang University, Hangzhou 310018, China.
Abstract:
This study established a method for determining trace anions in G2-grade phosphoric acid under ultra-clean conditions that used two-dimensional ion-exclusion-ion-exchange chromatography coupled with valve-switching technology. The optimal flow rate of ultrapure water through the ion-exclusion column in the first dimension was determined to be 0.5 mL/min through comparative experiments. The valve-switching window was optimized to minimize enriching PO43- on the anion-enrichment column and reduce interference from the phosphoric-acid matrix. The use of external water mode for the suppressor led to lower baseline noise, and optimizing the elution program of the second dimension enabled target trace anions to be well separated, thereby avoiding the risk of a false positive for NO3-. The anions of Cl-, Br-, NO3-, and SO42- exhibited good linearities within the 0.5-20 μg/kg, with correlation coefficients ≥0.999. The limits of detection (LODs) and quantification (LOQs) were 0.09-0.29 μg/kg and 0.29-0.97 μg/kg, respectively. The spiked recoveries of each anion at four spiked levels ranged from 91.7% to 103.6%, with RSDs of 0.1%-4.9%. The developed method demonstrated excellent stability through repeated injections, and low detection limits; it is highly precise and accurate, and meets the detection requirements for trace anions in G2-grade phosphoric acid. Additionally, retention-time comparisons with single standard substances led to the discovery of other phosphorus-related anions and anion clusters, such as hypophosphite (HPO3-) and hexametaphosphate (PO3-)6, in G2-grade phosphoric acid, in addition to conventional anions. This finding is theoretically and practically significant for the future development of purification processes for higher-purity phosphoric acid.
More Related Videos
Related Concept Videos
Ion-Exchange Chromatography
Ion Exchange
Gas Chromatography: Types of Detectors-II
Size-Exclusion Chromatography
Silica particles offer advantages such as rigidity,...
Two-dimensional Gel Electrophoresis
The first dimension separation uses the isoelectric focusing or IEF technique performed on immobilized pH gradient (IPG) strips that separate proteins according to their isoelectric points.
Biological samples, such...
Gas Chromatography: Types of Columns and Stationary Phases
For an analyte to remain on the column for a sufficient amount of time, it must exhibit some level of compatibility (or...