Full Validation of an HPLC-UV Analytical Method for Azithromycin Quantification Using Comparative Approaches: Total
Wafaa El-Ghaly1,2, Asmae Elouari1, Lamia Zaari Lambarki3
1Mohammed VI University of Sciences and Health (UM6SS), Faculty of Pharmacy, Boulevard Mohammed Taïeb Naciri, Commune Hay Hassani, 82403, Casablanca, Morocco.
Background:
Azithromycin is a complex molecule derived from erythromycin. The control of its dosage in conventional release tablets requires the analytical validation of its method to ensure accurate quantification and provide confidence in the reliability of the results for informed decision-making.
Objective:
This study aims to validate an innovative method for azithromycin quantification using the accuracy profile. Additionally, a comparison is made between the uncertainty measurements calculated from the validation data using two formulas proposed by Feinberg et al. and Saffaj and Ihssane and contrasted with the ISO GUM approach.
Methods:
A liquid chromatography system intended for azithromycin analysis equipped with a reversed-phase C18 stationary phase consisting of octadecyl silyl vinyl polymer in a UV detector operating at 210 nm at a temperature of 40°C in isocratic elution using a mobile phase of acetonitrile and dipotassium hydrogen phosphate buffer (6.7 g/L), in the fraction of (60:40, v/v) at pH = 8.
Results:
The various accuracy profiles are illustrated to ensure that a known quantity of anticipated findings acquired through the method stand inside the tolerance interval of 95% and remain within the previously set acceptance limits of ±5%. Measurement uncertainty provides comparable values using both formulas of the total error approach. However, it was observed that the ISO-GUM approach tends to overestimate the expanded uncertainty. Specifically, while the ISO-GUM approach provides a rigorous framework, the use of the validation data offers a more empirical uncertainty estimation.
Conclusion:
The approach based on the total error grants the ability to accurately close the routine uncertainty, emphasizing a complete validation.
Highlights:
The proposed method is robust for pharmaceutical application, demonstrating good accuracy, with 95% of tolerance and uncertainty limits falling within the predefined acceptance limits of ±5%.