Comparative Performance Analysis of Manual and Updated Optional Calibration Algorithms for the CareSens Air CGM
Nina Jendrike1, Manuel Eichenlaub1, Manuela Link1
1Institut für Diabetes-Technologie, Forschungs- und Entwicklungsgesellschaft mbH an der Universität Ulm, Ulm, Germany.
Background:
The CE-marked CareSens Air continuous glucose monitoring (CGM) system (CSAir) features a 15-day sensor lifetime, a 2-hour warm-up period and mandatory manual calibrations. During subsequent product development, the algorithm was updated to reduce the warm-up period to 30 minutes and make user-entered calibrations optional. This study compared the CSAir's performance between the manual and updated algorithms.
Methods:
Thirty adults with diabetes wore three CSAir sensors on their upper arms for 15 days. The study included three in-clinic sessions with capillary comparator measurements at 15-minute intervals over seven hours and glucose manipulation in the hypo- or hyperglycemic range. Point accuracy was assessed via mean absolute relative difference (MARD), 20/20 agreement rates (AR) stratified by BG range, and sensor wear time. Further evaluations included clinical point accuracy, alert reliability, technical reliability, safety and user satisfaction.
Results:
The CSAir's updated algorithm exhibited improved accuracy compared with the manual calibration algorithm, with a total 20/20 AR of 93.9% (vs 90.1%) and an MARD of 8.7% (vs 9.9%). Accuracy remained stable across measurement ranges and sensor lifetime. Diabetes Technology Society Error Grid analysis revealed high clinical accuracy, with 88.0% and 92.4% of data pairs in zone A for the manual and updated algorithms, respectively. The estimated survival probability was 88.8%. Participants reported positive user satisfaction. No safety concerns were identified.
Conclusions:
Both algorithms of CSAir demonstrated robust performance and reliability with improved accuracy with the updated version. The study results of the CSAir suggest its suitability for nonadjunctive use.
Related Concept Videos
Instrument Calibration
Analytical Balance Calibration
An analytical balance measures mass and requires regular calibration to...
Errors occurring during blood pressure monitoring
Several factors...
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...