Visualization of Low-Level Gamma Radiation Sources Using a Low-Cost, High-Sensitivity, Omnidirectional Compton Camera
Published on: January 30, 2020
Gamma-ray spectra and absorbed doses measured at EuXFEL undulator system
Olga Falowska-Pietrzak1, Anders Hedqvist1, Fredrik Hellberg1
1Department of Physics, Stockholm University, Stockholm, Sweden.
None:
Characterization of the stray radiation field present in the undulator systems at the European XFEL GmbH (EuXFEL) is of great importance, as the potential damage to undulator permanent magnets, electronics and diagnostic equipment depends both on the type of particles and the energy. This work presents the energy profile of the stray radiation measured in the upstream and downstream part of the undulator system near the beam pipe. The influence of machine operation settings on the radiation field intensity was also investigated. A comparison between gamma-ray energy measurements (30 keV to 1.5 MeV) and Geant4 Monte Carlo simulations shows that stray radiation in the upstream part of the undulator system originates from high-energy electrons intersecting the beam pipe wall. The intensity of the measured signal is approximately linear to the charge passing through the undulator system and to the electron energy but is inversely proportional to the square root of the undulator gap. Measurements in the upstream part of the undulator system show that the stray radiation level can be mitigated with careful choice of accelerator settings. The intensity of the radiation with energies below approximately 400 keV is higher in the downstream part of the undulator system compared with the upstream part, during regular operation. A CZT spectrometer and RADFET measurements confirm that the radiation field in the downstream part of the undulator system is dominated by low-energy synchrotron radiation (below 200 keV), which constitutes approximately 99% of the stray radiation field.
Related Concept Videos
Biological Effects of Radiation
Atomic Absorption Spectroscopy: Radiation and Light Sources
Two common narrow-range 'line' sources used in AAS are hollow-cathode lamps (HCLs) and...
Atomic Emission Spectroscopy: Instrumentation
Atomic Emission Spectroscopy: Lab
Absorption of Radiation
Inductively Coupled Plasma Atomic Emission Spectroscopy: Instrumentation
There are three main types of inductively coupled plasma atomic emission spectroscopy (ICP-AES) instruments: sequential, simultaneous multichannel, and Fourier transform instruments, with the latter being less commonly used....