Validation of a Manual Method for Measuring Left Atrial Reservoir Strain Against Automated Speckle Tracking Analysis
Marina Leitman1,2, Vladimir Tyomkin1
1Department of Cardiology, Shamir Medical Center, Zerifin 70300, Israel.
Abstract:
Background: Left atrial strain, particularly reservoir strain, has emerged as a sensitive marker of left atrial function and an early indicator of diastolic dysfunction and cardiovascular risk. However, automated left atrial strain analysis is not universally available, particularly in resource-limited settings. In this study, we propose a manual method for calculating biplane left atrial reservoir strain and validate its agreement with automated software in patients with atrial fibrillation and in sinus rhythm. Methods: Echocardiography examinations from 30 patients with atrial fibrillation and 30 patients in sinus rhythm were analyzed. Left atrial reservoir strain was calculated using both an automatic speckle tracking imaging-based algorithm and a manual point-by-point method based on atrial wall delineation. Agreement between methods was assessed using Pearson correlation, Bland-Altman analysis, and intraclass correlation coefficient. Results: Strong correlation and excellent agreement were observed between the two methods in both groups. Pearson correlation coefficients were r = 0.95 (p < 0.0001) in the atrial fibrillation group and r = 0.94 (p < 0.0001) in the sinus rhythm group. Bland-Altman analysis showed narrow limits of agreement, particularly in the atrial fibrillation group. The intraclass correlation coefficient was 0.95 in atrial fibrillation and 0.92 in sinus rhythm, indicating excellent reliability. The standard error of measurement and minimal detectable change were low in both groups. Conclusions: Manual measurement of left atrial reservoir strain is feasible, reproducible, and demonstrates excellent agreement with automated software. It may serve as a reliable alternative in clinical scenarios where automated tools are unavailable.
More Related Videos
09:05Transthoracic Speckle Tracking Echocardiography for the Quantitative Assessment of Left Ventricular Myocardial Deformation
Published on: October 20, 2016
08:10Estimating Bilateral Atrial Function by Cardiovascular Magnetic Resonance Feature Tracking in Patients with Paroxysmal Atrial Fibrillation
Published on: July 20, 2022
Related Concept Videos
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Assessment of blood pressure in brachial artery(one-step method)
Prepare for the Procedure:
Assessment of blood pressure in brachial artery(two-step method)
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment:
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.